Visualase Thermal Therapy System, Software: Visualase Software versions 3.1.1 - 3.2, Product #: 9735542
Recall
- Recall Number
- Z-3251-2018
- Event Number
- 80308
- Firm
- Medtronic Navigation, Inc.
- FEI Number
- 1000517638
- Product Code
- LLZ
- Status
- Terminated
- Root Cause
- Under Investigation by firm
- Initiated
- June 1, 2018
- Terminated
- February 2, 2022
- Address
- 826 Coal Creek Cir, Louisville, CO, 80027-9710
Description
Visualase Thermal Therapy System, Software: Visualase Software versions 3.1.1 - 3.2, Product #: 9735542
Inaccuracy of MR thermometry during MRI-guided laser ablation procedures using the Medtronic Visualase Thermal Therapy System may result in unaccounted spread of thermal energy to the surrounding tissue.
On 06/01/2018, Urgent Medical Device Correction letters and Consignee Response Forms were mailed to customers via overnight mail service. Customers were provided with additional information intended to supplement the system manual and instructed to call the firm at 1-888-826-5603 if they had any questions.
U.S.: MN, NY, PA, FL, AZ, MA, VA, CA, CO, AL, WI, TX, MO, DC, OH, NH, GA, SC, NJ, WA, CT, MI, OK, IN, MD, IL, KY, LA, OR, RI, UT, AR, MS, TN, NC Foreign: United Kingdom, Germany, Switzerland, Netherlands, Israel
161