FDA Recall Open, Classified

LINK Universal Handle, with quick coupling, Stainless Steel, straight, Item Number 130-394/01

Recall: Z-3245-2024 · Initiated September 11, 2024

Recall

Recall Number
Z-3245-2024
Event Number
95344
Firm
Waldemar Link GmbH & Co. KG (Corp. Hq.) Barkhausenweg 10 Hamburg Germany
FEI Number
3002808472
Product Code
LXH
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
September 11, 2024
Posted
September 24, 2024

Description

LINK Universal Handle, with quick coupling, Stainless Steel, straight, Item Number 130-394/01

Reason

Some lots of the LINK Universal Handle, with quick coupling, Stainless Steel, straight may suffer an immediate material failure intraoperatively when the instrument is used for the first time. The instrument can break in the middle.

Action

An URGENT: MEDICAL DEVICE RECALL notification letter dated 9/11/24 was sent to customers. Actions to be taken by the Customer/User: Please check your inventory for this item and lot combination and return any affected product to LinkBio. Please respond using the attached Distributor Reply form to indicate your understanding of this Device Recall and of the need to return any affected product. Pre-paid shipping labels will be provided once the Distributor Reply Form is returned, should any affected product be identified. Please return any affected product as soon as possible to the following address: LinkBio Corporation 69 King St Dover, NJ 07801 Please contact LinkBio at 973-625-1333 or [email protected] for any questions.

Distribution

US Nationwide distribution in the states of FL, KS, MI, NJ, VA.

Quantity

32 devices