FDA Recall Terminated

DX-D 600 Product Usage: The DX-D 600 system is a ceiling mounted General Radiography X-Ray imaging system used in hospitals, clinics and medical practices by physicists, radiographers and radiologists to make, process and view static X-Ray radiographic images of the skeleton (including skull, spinal column and extremities), chest, abdomen and other body parts on adult, pediatric or neonatal patients.

Recall: Z-3245-2018 · Initiated August 15, 2018

Recall

Recall Number
Z-3245-2018
Event Number
81046
Firm
Agfa N.V. Septestraat 27 Mortsel Belgium
FEI Number
3001556265
Product Code
KPR
Status
Terminated
Root Cause
Component design/selection
Initiated
August 15, 2018
Posted
September 25, 2018
Terminated
June 23, 2020

Description

DX-D 600 Product Usage: The DX-D 600 system is a ceiling mounted General Radiography X-Ray imaging system used in hospitals, clinics and medical practices by physicists, radiographers and radiologists to make, process and view static X-Ray radiographic images of the skeleton (including skull, spinal column and extremities), chest, abdomen and other body parts on adult, pediatric or neonatal patients.

Reason

A supplier for the Agfa system noted a potential for the steel support cables of a similar device to break. This led to a mandatory review of all systems, changing the pulley and cables if any deterioration in the cables or irregular or excessive pulley wear. If the two cables that support the equipment to its roof anchor break, the equipment might fall and cause harm to the patient, or others.

Action

On August 15, 2018, an "URGENT FIELD SAFETY NOTICE" letter was emailed to the US and Canadian consignees. The letter described the product issue: Breakage of the steel cables that support the equipment to its roof anchor may cause the equipment to fall down and potentially harm the patient, user, or third parties. Actions undertaken by Agfa NV: Soon Agfa NV or its representative will visit your site to do a verification/inspection of your unit related to this topic. Despite the fact that this issue potentially affects older systems with high workload, Agfa NV or its representative will verify/inspect all DX-D 600 systems. Recommended actions to be taken by the customer: If you notice an abnormal displacement in the vertical axis or any suspicious noise during the vertical movement, please notify your Agfa NV Service contact at once. Customers were asked to complete the attached feedback form to confirm receipt.

Distribution

Worldwide Distribution - US Nationwide in the states of CA, DC, FL, IL, IN, KS, LA, MA, MN, MO, NV, NY, OH, OR, SC and TN. Foreign distribution to Canada.

Quantity

108