FDA Recall Open, Classified

ACUSON Redwood 1.0 and 2.0 Diagnostic Ultrasound Systems, REF: 11503314, with software

Recall: Z-3216-2024 · Initiated August 15, 2024

Recall

Recall Number
Z-3216-2024
Event Number
95254
Firm
Siemens Medical Solutions USA, Inc.
FEI Number
3023245
Product Code
IYN
Status
Open, Classified
Root Cause
Software design
Initiated
August 15, 2024
Posted
September 24, 2024
Address
22010 Se 51st St, Issaquah, WA, 98029-7298

Description

ACUSON Redwood 1.0 and 2.0 Diagnostic Ultrasound Systems, REF: 11503314, with software

Reason

If ultrasound systems with software, are changed from factory default to : 1) Milliliters per second (ml/sec, mL/sec) or 2) Milliliters per minute (ml/min, mL/min); then systems will perform incorrect unit conversion of liters to milliliters using multiplier 100 instead of 1000, resulting values 10x smaller than actual, which could contribute misdiagnosis or negatively influence patient management

Action

On 8/15/2024, correction notices were distributed to customers who were asked to do the following: To avoid potential misdiagnosis by an under-represented volume flow measurement value, use the factory default display configuration of l/min (liters per minute). Review of previous examination results ONLY if a clinical assessment reviewed a volume flow measurement(s) with the unit of measurement configured to display a value in milliliters. Firm's Customer Service Engineer will contact you to schedule a facility visit to update the system or inform you of a remote update when the software update is available. The software update is currently under development and estimated to be available in the summer of 2024. Ensure that all users of ACUSON Juniper and ACUSON Maple systems within your organization, and others who may need to be informed, receive the safety relevant information provided with this notice and take the actions specified herein. If you have further questions, please contact the firm's Ultrasound Service Customer Care Center at 1-800-888-7436.

Distribution

US: MS, WA, TX, NJ, GA, PN, NY, AL, MN, NE, IN, CT, CA, IL, TN, WI, ND, NM, AR, NC, PR, RI, WY, MD, FL, HI, AZ, SC, MO, OK, VA, ME, KS, MI, LA, IA, OH, SD, NH, KY, AL, OR, MA, GU. OUS: Afghanistan, Albania, Algeria, Andorra, Angola, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia-Herzegovina, Botswana, Brazil, Brunei, Bulgaria, Cameroon, Canada, Chile, China, Colombia, Congo, the Democratic Republic of the, Costa Rica, C¿te dIvoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Eritrea, Ethiopia, Faroe Islands, Fiji, Finland, France, French Guinea, French Polynesia, Georgia, Germany, Greece, Guatemala, Guernsey, Guyana, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Islamic Republic of, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kosovo, Kuwait, Latvia, Lebanon, Libya, Liechtenstein LI B, Lithuania, Luxembourg, Macedonia, Malawi, Malaysia, Malta, Mexico, Moldova, Rep. of, Mongolia, Montenegro, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Republic Korea, Reunion, Romania, Ruanda, Russian Federation, San Marino, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Swaziland, Sweden, Switzerland, Taiwan, Tajikistan, Tanzania, Thailand, Trinidad,Tobago, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Vietnam, Yemen, Zambia, Zimbabwe

Quantity

1990