FDA Recall Open, Classified

Percussionaire, Inc., REF: P5-TEE-20, Qty: 20, Rx Only. IPV (Intrapulmonary Percussive Ventilation) therapy In-Line Valve

Recall: Z-3208-2024 · Initiated August 21, 2024

Recall

Recall Number
Z-3208-2024
Event Number
95239
Firm
Percussionaire Corporation
FEI Number
1000524541
Product Code
BZO
Status
Open, Classified
Root Cause
Device Design
Initiated
August 21, 2024
Posted
September 27, 2024
Address
130 Mcghee Rd, Ste 109, Sandpoint, ID, 83864-8409

Description

Percussionaire, Inc., REF: P5-TEE-20, Qty: 20, Rx Only. IPV (Intrapulmonary Percussive Ventilation) therapy In-Line Valve

Reason

Updated labeling to ensure that blue cap on in-line valve is removed, prior to delivery of (Intrapulmonary percussive ventilation) IPV therapy.

Action

On September 3, 2024, Percussionaire Inc. issued a "Urgent Medical Device Field Safety Notice" to affected consignees. Percussionaire asked consignees to take the following actions: 1. Complete and Return Acknowledgement form (see Appendix 1) after reviewing and implementing the requested actions below by September 27, 2024. 2. Review updated instructions for proper in-line use (see Appendix 3) and updated IFU (see Appendix 4) 3. Print and post updated instructions for proper In-Line Valve use (Appendix 3) throughout your facility for adequate distribution of new information to users. 4. Communicate and train with newly presented instructions for proper use to primary users of IPV therapy, specifically for in-line use. 5. Review institutional protocols related to P5-TEE inline use and update as needed to conform with the updated instructions. 6. Re-label current product by placing the sticker provided by Percussionaire on each individual P5-TEE bag on stock (stickers to be provided via mail). 7. Report any adverse events to [email protected] and/or to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail or by fax. 8. Report any quality problems experienced with the use of this product to Percussionaire/Sentec Customer Service department via email to [email protected].

Distribution

US: Nationwide OUS: Israel Japan Russia Switzerland Canada Turkey

Quantity

2523 units