400mL, Premium, Urine Meter, 16 FR, Silicone Temperature Sensing Foley Traywith Cath-Secure Plus, Sterile, Item Code P4P16TSD
Recall
- Recall Number
- Z-3206-2018
- Event Number
- 80747
- Firm
- COVIDIEN LLC
- FEI Number
- 1282497
- Product Code
- KOD
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- July 11, 2018
- Terminated
- August 3, 2020
- Address
- 15 Hampshire St, Mansfield, MA, 02048-1113
Description
400mL, Premium, Urine Meter, 16 FR, Silicone Temperature Sensing Foley Traywith Cath-Secure Plus, Sterile, Item Code P4P16TSD
The temperature sensor catheters may be defective in that they will show a lower body temperature measurement than the actual temperature of the patient.
On 7/11/18, an Urgent Medical Device Recall letter was distributed to customers. The letter instructed customers to do the following: 1. Please quarantine any unused product of the Covidien item codes and lot numbers listed above (see Attachment A to locate). 2. If you have distributed any of the affected Covidien catheters or kits, then please promptly forward a copy of this letter to those recipients. 3. All unused product of affected item codes and lot numbers must be returned. 4. Please return unused affected product as explained below. You must complete the Recalled Product Return Form (Attachment B) regardless of whether or not you have inventory of affected product.
The products were distributed to the following US states: CA, CO, FL, GA, IA, IL, IN, KY, LA, MA, MD, MO, NC, NJ, NY, OH, OR, PA, RI, SC, TN, TX, VA, and WA.
27515 total