FDA Recall Terminated

400mL, Premium, Urine Meter, 14 FR, Silicone Temperature Sensing Foley Traywith Cath-Secure Plus, Sterile, Item Code P4P14TSD

Recall: Z-3205-2018 · Initiated July 11, 2018

Recall

Recall Number
Z-3205-2018
Event Number
80747
Firm
COVIDIEN LLC
FEI Number
1282497
Product Code
KOD
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
July 11, 2018
Terminated
August 3, 2020
Address
15 Hampshire St, Mansfield, MA, 02048-1113

Description

400mL, Premium, Urine Meter, 14 FR, Silicone Temperature Sensing Foley Traywith Cath-Secure Plus, Sterile, Item Code P4P14TSD

Reason

The temperature sensor catheters may be defective in that they will show a lower body temperature measurement than the actual temperature of the patient.

Action

On 7/11/18, an Urgent Medical Device Recall letter was distributed to customers. The letter instructed customers to do the following: 1. Please quarantine any unused product of the Covidien item codes and lot numbers listed above (see Attachment A to locate). 2. If you have distributed any of the affected Covidien catheters or kits, then please promptly forward a copy of this letter to those recipients. 3. All unused product of affected item codes and lot numbers must be returned. 4. Please return unused affected product as explained below. You must complete the Recalled Product Return Form (Attachment B) regardless of whether or not you have inventory of affected product.

Distribution

The products were distributed to the following US states: CA, CO, FL, GA, IA, IL, IN, KY, LA, MA, MD, MO, NC, NJ, NY, OH, OR, PA, RI, SC, TN, TX, VA, and WA.

Quantity

27515 total