Xper Flex Cardio Physiomonitoring System Model # 453564669081 Software Version 1.6.0.0388
Recall
- Recall Number
- Z-3192-2018
- Event Number
- 80761
- Firm
- Invivo Corporation
- FEI Number
- 1051786
- Product Code
- MWI
- Status
- Terminated
- Root Cause
- Software Manufacturing/Software Deployment
- Initiated
- August 3, 2018
- Terminated
- July 15, 2021
- Address
- 12151 Research Pkwy, Ste 200, Orlando, FL, 32826-3286
Description
Xper Flex Cardio Physiomonitoring System Model # 453564669081 Software Version 1.6.0.0388
Released with an incorrect, older firmware version installed. May result in intermittent loss of ECG monitoring, inaccurate ECG amplitude display and or auxiliary output, and inaccurate heart rate (HR) displayed due to QRS detection fault.
Philips is asking customers to follow the Action to be Taken by Customer/User section of the August 3, 2018, Field Safety Notice: The Field Safety Notice informs customers of the issue, identifies details of the units affected, gives instructions on actions to be taken by the customer and identifies what action Philips plans to take to remedy the issue. Philips will replace all affected devices. Users should discontinue use of affected devices immediately. A Philips representative will contact you regarding your affected device.
Distribution in the Netherlands
38