FDA Recall Open, Classified

Creatine Kinase (CK) used on cobas c 311/501/502 and COBAS INTEGRA 400 plus-In vitro test for the quantitative determination of creatine kinase (CK) in human serum and plasma. Catalog Number: 07190794190

Recall: Z-3187-2024 · Initiated August 15, 2024

Recall

Recall Number
Z-3187-2024
Event Number
95221
Firm
Roche Diagnostics Operations, Inc.
FEI Number
1823260
Product Code
JHS
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
August 15, 2024
Posted
September 20, 2024
Address
9115 Hague Rd, Indianapolis, IN, 46256-1025

Description

Creatine Kinase (CK) used on cobas c 311/501/502 and COBAS INTEGRA 400 plus-In vitro test for the quantitative determination of creatine kinase (CK) in human serum and plasma. Catalog Number: 07190794190

Reason

Creatine Kinase (CK) Assay Abnormal calibrations on the cobas c 311/501/502 and COBAS INTEGRA 400 plus Analyzers may lead to erroneous results and misdiagnosis.

Action

Roche issued Urgent Medical Device Corrections,TP-02225 and TP-02227 via UPS ground (08/15/2024). Roche engaged Sedgwick a third party recall company. Letter states reason for recall, health risk and action to take: Creatine Kinase kits, lot number 755471 expired on 06/30/2024. Ensure any inventory you have of this lot has been discarded per your local guidelines. " Follow your laboratory s standard operating procedures to determine specific clinical implications for your patients and whether retrospective review of patient results is required. " Consult with the physician or pathologist at your facility to determine specific clinical implications for your patients. " If your facility has distributed the affected product to another site, please ensure this UMDC is provided to that site. " If you received Recall Response Form (TP-02226) with this UMDC, complete all sections and fax it to 1-800-898-6169 or email it to [email protected] by September 1, 2024. If you did not receive the Recall Response Form with this UMDC, disregard this action. " File this UMDC for future reference. Please contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336 if you have questions about the information in this UMDC.

Distribution

US Nationwide distribution.

Quantity

6504 units