FDA Recall Open, Classified

Beckman Coulter DxI 9000 Access Immunoassay Analyzer, Model Number C21510, RSM Pneumatic Tubing Kit item #2, REF C11137; in vitro diagnostic instrument

Recall: Z-3180-2024 · Initiated July 10, 2024

Recall

Recall Number
Z-3180-2024
Event Number
95156
Firm
Beckman Coulter, Inc.
FEI Number
2122870
Product Code
JJE
Status
Open, Classified
Root Cause
Nonconforming Material/Component
Initiated
July 10, 2024
Posted
September 19, 2024
Address
1000 Lake Hazeltine Dr, Chaska, MN, 55318-1037

Description

Beckman Coulter DxI 9000 Access Immunoassay Analyzer, Model Number C21510, RSM Pneumatic Tubing Kit item #2, REF C11137; in vitro diagnostic instrument

Reason

Beckman Coulter is recalling the DxI 9000 Access Immunoassay Analyzer due to undersized Pneumatics Tubing in the reagent storage module causing air leaks, compressor failures or sample processing errors. Four complaints were filed for this event that led to Beckman Coulter finding undersized tubing in stock and initiated corrective actions. Use of the affected device may result in errors leading to delay of results.

Action

Beckman Coulter issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 7/10/2024 via US mail and email. The notice explained the problem, impact, and requested the following actions be taken: If your laboratory observes either of these system eventscontact the Beckman Coulter Customer Technical Support Center at 1-800-854-3633 (US and Canada) or your local Beckman Coulter representative (outside US and Canada: a). Chassis.Pneumatics MachineControlStatus(Chassis maintain system air pressure failed. 1.The compressor has failed. 2.A large leak exists in the system) b). Pneumatics Air pressure under limit: ### PSI, Chassis.Pneumatics. WS04CompressedAir.32-35psi.PressureSensor. Beckman Coulter will schedule a time to replace the RSM pneumatic tubing.

Distribution

Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Belgium, Brazil, Croatia, Czech Republic, Egypt, France, Germany, Hungary, Ireland, Israel, Italy, Netherlands, New Zealand, Poland, Portugal, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, United Kingdom.

Quantity

170 units