FDA Recall Open, Classified

System Name: 708038 MultiDiagnost Eleva with Flat Detector 708037 MultiDiagnost Eleva with Flat Detector 708036 MultiDiagnost Eleva 708034 MultiDiagnost Eleva with Flat Detector 708032 MultiDiagnost Eleva

Recall: Z-3168-2024 · Initiated August 22, 2024

Recall

Recall Number
Z-3168-2024
Event Number
95256
Firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Veenpluis 4-6 Best Netherlands
FEI Number
3002807880
Product Code
KPR
Status
Open, Classified
Root Cause
Device Design
Initiated
August 22, 2024
Posted
September 18, 2024

Description

System Name: 708038 MultiDiagnost Eleva with Flat Detector 708037 MultiDiagnost Eleva with Flat Detector 708036 MultiDiagnost Eleva 708034 MultiDiagnost Eleva with Flat Detector 708032 MultiDiagnost Eleva

Reason

Potential Detachment of the Cable Hose Carrier (due to broken plastic and/or bolts) which May Result in Parts Falling and/or Part of the Cable Hose Dropping, and result in injury.

Action

Philips Issued URGENT Medical Device Correction Letter (Philips Reference # 2023-IGT-BST-006) on 8/22/24. Letter states reason for recall, health risk and action to take: a. Affected systems may continue to be used in accordance with their Instructions for Use (IFU) and the instructions below. b. Avoid that operators or bystanders stand underneath the Cable Hose Carriers, especially when moving or turning the system monitor. c. Avoid having the patient located underneath the Cable Hose Carriers (e.g., during transport). d. If you notice loose or broken bolts and/or broken plastic parts of the Cable Hose Carriers (see Figure 2), please contact your local Philips representative immediately. e. As part of the preventive maintenance cycle, Philips will check the Monitor Ceiling Suspension, including the Cable Hose Carriers, as indicated in the Preventive Maintenance Manual Update attached in Appendix B. o Keep a copy of this Preventative Maintenance Manual Update with your current manual. o If you do not use Philips to perform the preventative maintenance on your system, provide a copy of the Preventive Maintenance Manual Update to your qualified and authorized service provider. f. Circulate this Urgent Medical Device Correction letter to all users of the system so that they are aware of the issue and follow the instructions above. Keep this Medical Device Correction letter with the documentation of the system until Philips corrects your system. g. Complete and return the attached response form (page 4) to Philips promptly and no later than 30 days from receipt. Completing this form confirms receipt of the Urgent Medical Device Correction letter and understanding of the issue and required actions to be taken. Philips is redesigning the Cable Hose Carriers. Once the redesigned cable hose carriers are available, Philips will contact you to schedule a visit to install them in the affected system(s) (reference FCO72200536, FCO72200576, and FCO72200577). As of the date of t

Distribution

Nationwide, Canada and Worldwide: Albania Algeria Argentina Armenia Australia Austria Azerbaijan Bahrain Bangladesh Belarus Belgium Bolivia Bosnia and Herzegovina Brazil Brunei Darussalam Bulgaria Cambodia Cayman Islands Chile China Colombia Costa Rica C¿te D'Ivoire Croatia Cuba Curacao Cyprus Czech Republic Denmark Dominican Republic Ecuador Egypt Estonia Ethiopia Fiji Finland France French Polynesia Georgia Germany Ghana Gibraltar Greece Guadeloupe Guinea Honduras Hong Kong Hungary Iceland India Indonesia Iran Iraq Ireland Israel Italy Jamaica Japan Jordan Kazakhstan Kenya Korea, Republic of Kosovo Kuwait Kyrgyzstan Lao People's Democratic Republic Latvia Lebanon Libya Lithuania Macao Macedonia Malaysia Maldives Malta Martinique Mauritania Mauritius Mexico Mongolia Morocco Myanmar Nepal Netherlands New Caledonia New Zealand Nicaragua Norway Oman Pakistan Palestine, State of Panama Papua New Guinea Paraguay Peru Philippines Poland Portugal Qatar R¿union Romania Russian Federation Saudi Arabia Senegal Serbia Singapore Slovakia Slovenia South Africa Spain Sri Lanka Sudan Sweden Switzerland Syrian Arab Republic Taiwan Tanzania Thailand Trinidad and Tobago Tunisia Turkey Turkmenistan Ukraine United Arab Emirates United Kingdom Uruguay Uzbekistan Venezuela Viet Nam Virgin Islands (U.S.) Yemen

Quantity

109 units