FDA Recall Open, Classified

Ivenix Infusion System (IIS), LVP Software; LVP-SW-0004, Software Version 5.9.1 and prior. Infusion pump software for infusion management system.

Recall: Z-3153-2024 · Initiated August 30, 2024

Recall

Recall Number
Z-3153-2024
Event Number
95240
Firm
Fresenius Kabi USA, LLC
FEI Number
3014732157
Product Code
FRN
Status
Open, Classified
Root Cause
Software design
Initiated
August 30, 2024
Posted
September 26, 2024
Address
50 High St, Ste 50, North Andover, MA, 01845-2620

Description

Ivenix Infusion System (IIS), LVP Software; LVP-SW-0004, Software Version 5.9.1 and prior. Infusion pump software for infusion management system.

Reason

The software has anomalies that have the potential to cause alarms, nonfunctioning pump, or unresponsive screen while continuing therapy. These could cause serious patient harm or death.

Action

An URGENT Medical Device Field Correction Important Safety Update notification letter dated 8/30/24 was sent to customers. Required Actions for Users: 1. Please fill out the Customer Reply Form attached to this letter to acknowledge receipt. 2. If your facility further distributes or transfers products to satellite sites or other locations, please ask these locations to contact the email address on the Customer Reply Form so we can send notice to those satellites directly. 3. Install new software version (5.9.2) for the Ivenix Large Volume Pump (LVP). See details below. How To Obtain and Install Software Version 5.9.2 To request installation, please reach out promptly to your Fresenius Kabi representative (1-855- 354-6387 or [email protected]). They will push the new software update to your pumps. When the LVP version 5.9.2 software update has been sent to the LVP, the following prompt appears before the LVP is shut down. Customer support - email Ivenix_su [email protected] Customer support - phone 1-855-354-6387 Normal business Hours are Monday through Friday 8:30 AM to 5 PM Eastern Time, After Hours Supported with Voicemail Messaging

Distribution

US Nationwide.

Quantity

17 units