FDA Recall Open, Classified

vanSonnenberg Sump Locking Pigtail Regular Kit Sump Catheter System Kit, Material Number REF M001273090; for use as a percutaneously placed drain for intra-abdominal fluid collections.

Recall: Z-3150-2024 · Initiated July 18, 2024

Recall

Recall Number
Z-3150-2024
Event Number
95186
Firm
Boston Scientific Corporation
FEI Number
3002095335
Product Code
FFA
Status
Open, Classified
Root Cause
Packaging process control
Initiated
July 18, 2024
Posted
September 13, 2024
Address
1 Scimed Pl, Maple Grove, MN, 55311-1565

Description

vanSonnenberg Sump Locking Pigtail Regular Kit Sump Catheter System Kit, Material Number REF M001273090; for use as a percutaneously placed drain for intra-abdominal fluid collections.

Reason

There is the potential for holes in the clear sterile barrier of the device pouch which may impact device sterility.

Action

Boston Scientific issued am Urgent Medical Device Removal notice to it consignees on 07/18/2024 via overnight mail. The notice explained the issue, potential hazard, and requested the return of the devices. Consignees were directed to post the notice in a visible location near the affected products to ensure visibility and to share the notice with all those to whom the device was transferred. Distributors were direct to notify their customers.

Distribution

US, Colombia, Philippines, Russia

Quantity

4 units