FDA Recall Open, Classified

AET36 ANESTH EXT SET - Anesthesia Extension Set - Normally Closed Dual Manifold, Four-way Stopcock, Model/Catalog Number: 472036

Recall: Z-3144-2024 · Initiated August 8, 2024

Recall

Recall Number
Z-3144-2024
Event Number
95153
Firm
B. Braun Medical, Inc.
FEI Number
2523676
Product Code
FPA
Status
Open, Classified
Root Cause
Process change control
Initiated
August 8, 2024
Posted
September 13, 2024
Address
901 Marcon Blvd, Allentown, PA, 18109-9512

Description

AET36 ANESTH EXT SET - Anesthesia Extension Set - Normally Closed Dual Manifold, Four-way Stopcock, Model/Catalog Number: 472036

Reason

Potential for product leakage from the Spin Lock Connector on the distal end of the device due to the migration of excess solvent during the assembly process.

Action

An URGENT MEDICAL DEVICE RECALL NOTIFICATION letter dated 8/6/24 was sent to customers. Actions Required by B. Braun Medical Inc. (BBMI) Customer/User: 1. Review the Medical Device Recall Notification in its entirety and ensure that all users in your organization of the above-mentioned product, and other concerned persons are informed about this voluntary recall. If you are a distributor and have further distributed the product, please forward this notice to your consignees. The recall is to be extended to the consumer level. 2. Determine your current inventory of the affected items within inventory of your facility, cease use and quarantine product subject to recall. Do not destroy any affected product. 3. Utilizing the attached "Medical Device Recall Acknowledgement Form", record the total number of individual impacted units. If you have no inventory remaining, please enter zero (0) on the form. 4. Return the completed "Medical Device Recall Acknowledgement Form" to B. Braun Medical Inc. Quality Assurance department by faxing the form to (610) 849-1197 or e-mail to [email protected] within two (2) weeks of receipt, even if the total inventory in your possession is zero (0). 5. Once we receive your Medical Device Recall Acknowledgement Form, a B. Braun Customer Support representative will contact you with instructions on how to return any impacted cases, including partial cases, in your possession and provide credit and/or replacement of the product based on your individual needs. Adverse reactions or quality problems experienced with this product, or questions about this recall may be reported to BBMl's Postmarket Surveillance Department by calling 1-844-903-6417. They may also be reported to the FDA's MedWatch Adverse Event Reporting program either online, by regular mail or by fax.

Distribution

US Nationwide distribution.

Quantity

19050 units