FDA Recall Open, Classified

Alaris Pump Module Model 8100, a modular infusion pump and monitoring system

Recall: Z-3142-2024 · Initiated August 15, 2024

Recall

Recall Number
Z-3142-2024
Event Number
94937
Firm
CareFusion 303, Inc.
FEI Number
2016493
Product Code
FRN
Status
Open, Classified
Root Cause
Process control
Initiated
August 15, 2024
Posted
September 13, 2024
Address
10020 Pacific Mesa Blvd, San Diego, CA, 92121-4386

Description

Alaris Pump Module Model 8100, a modular infusion pump and monitoring system

Reason

Due to affected devices that may have bent female IUI connector and may not be operable and potentially cause delays in the start of the infusion or monitoring.

Action

On 08/15/2024, the firm sent an "URGENT MEDICAL DEVICE PRODUCT CORRECTION" Letter via FedEx /2-day or email to customers informing them that, BD has identified an increase in bent contacts on the new Female/Left IUI connectors which may cause delays in initiating therapy. Customer are instructed to: 1. Per current labeling, customers should inspect all IUI connectors prior to use. You may also refer to Appendix A for further details. 2. If you identify a bent/lifted IUI pin prior to use, do not use the device and send the device to biomedical engineering or Qualified Service Personnel for repair. 3. Please distribute this notice to all facility personnel responsible for inspection of BD Alaris" devices. For questions/assistance - Technical Questions - Technical Support - Phone: 1-866-488-1408, Phone hours: 5:00am to 5:00pm PT, Monday Friday. Product Complaints - BD Designated Complaint Handling Unit - Phone: 1-888-812-3266 Phone hours: 7:00am to 5:00pm PT, Monday Friday, Email: [email protected]. Email for receipt of customer response form and recall related questions -

Distribution

Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV. The countries of United Arab Emirates, Australia, Belgium, Canada, Hungary, Kuwait, Qatar, Saudi Arabia, Singapore, Taiwan, and New Zealand.

Quantity

192964