FDA Recall Terminated

Spectra-System/Legacy Guided Surgery Handle Kit, Device Listing Number D245883. The Guided Handle Kit (G-HK) allows clinicians to drill through a surgical guide to create an osteotomy for implant placement with the precision and accuracy as designed with treatment planning software. The G-HK consist of 4 guide handles with inserts attached on each end. Each insert has an opening that matches the diameter of our surgical drills. The insert portion of the guide handle fits intimately into the guide sleeve that is embedded into a surgical guide.

Recall: Z-3115-2017 · Initiated June 26, 2017

Recall

Recall Number
Z-3115-2017
Event Number
77554
Firm
Implant Direct Sybron Manufacturing, LLC
FEI Number
3001617766
Product Code
LXH
Status
Terminated
Root Cause
Process control
Initiated
June 26, 2017
Terminated
June 16, 2021
Address
3050 E Hillcrest Dr, Westlake Village, CA, 91362-3171

Description

Spectra-System/Legacy Guided Surgery Handle Kit, Device Listing Number D245883. The Guided Handle Kit (G-HK) allows clinicians to drill through a surgical guide to create an osteotomy for implant placement with the precision and accuracy as designed with treatment planning software. The G-HK consist of 4 guide handles with inserts attached on each end. Each insert has an opening that matches the diameter of our surgical drills. The insert portion of the guide handle fits intimately into the guide sleeve that is embedded into a surgical guide.

Reason

The inside diameter of the 2.3mm small drill guide inserts have a smaller diameter than the specifications require.

Action

Implant Direct sent an Safety Alert /Advisory Notice letter dated June 26, 2017. Instructions included to review and remove inventory for return to Implant Direct, complete and return the acknowledgement and recall return form, and contact customers and notify the of the recall if product was further distributed. For further questions, please call (888) 649-6425.

Distribution

Worldwide Distribution - USA (nationwide) Distribution to the states of : AZ, CA, CO, CT, FL, ID, IL, KS, MI, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OR, PA, SC, TN, TX, UT, VA, WA., and to the countries of : Canada, Netherlands, Spain, Great Britain, Germany and Switzerland.

Quantity

84 units