Coulter HmX CP and Coulter HmX AL Analyzers The purpose of the HmX Hematology Analyzer is to separate the normal patient, with all normal system-generated parameters, from the patient who needs additional studies. These studies include further measurements of cell size and cell distribution, biochemical investigation or any other test that helps diagnose the abnormality
Recall
- Recall Number
- Z-3114-2017
- Event Number
- 77830
- Firm
- Beckman Coulter Inc.
- FEI Number
- 2050012
- Product Code
- GKZ
- Status
- Terminated
- Root Cause
- Labeling Change Control
- Initiated
- June 26, 2017
- Terminated
- September 14, 2023
- Address
- 250 S Kraemer Blvd, Brea, CA, 92821-6232
Description
Coulter HmX CP and Coulter HmX AL Analyzers The purpose of the HmX Hematology Analyzer is to separate the normal patient, with all normal system-generated parameters, from the patient who needs additional studies. These studies include further measurements of cell size and cell distribution, biochemical investigation or any other test that helps diagnose the abnormality
Beckman Coulter has determined that additional clarification for the Blast Suspect messages is necessary. In rare situations, the LH 750, LH 780, LH 500 and HmX Analyzers may not flag or detect blasts in some blood samples.
Beckman Coulter sent an Urgent Medical Device Recall letter dated June 30, 2017, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to share the information with their laboratory staff. If they forwarded any of the affected product to another laboratory, please provide them a copy of the letter. Customers with questions should contact the Customer Support Center at 800-528-7694.
Worldwide Distribution - US (nationwide) Internationally to Albania, Algeria, Andorra, Angola, Antigua and Barbuda, Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belgium, Bermuda, Bosnia and Herzegovina, Brazil, Bulgaria, Burundi, Canada, China, Colombia, Congo, Croatia, Czech republic, Ecuador, Egypt, Ethiopia, Finland, France, French Polynesia, Germany, Greece, Guadeloupe, Guatemala, Honduras, Hong Kong, Hungary, Iceland, india, Indonesia, Italy, Japan, Iraq, Jordan. Kenya, Kuwait, Libya, Lithuania, Macau, Lebanon, Macedonia, Malawi, Malaysia, Mexico, Mongolia, Morocco, Mozambique, Netherlands, New Zealand, Nigeria, Pakistan, Panama, Paraguay, Philippines, Portugal, Puerto Rico, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sri Lanka, Swaziland, Sweden, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Turkey, UAE, UK, Uruguay, Venezuela, Vietnam, Virgin Islands, Yemen, Zambia, Zimbabwe.
3326 units