FDA Recall Terminated

Coulter LH 750, Coulter LH 780, and Coulter LH 500 Analyzers Hematology Analyzers are quantitative, automated hematology analyzers and leukocyte differential counters For In Vitro Diagnostic used in clinical laboratories. They also provide automated Reticulocyte analysis and enumeration of nucleated red blood cells (NRBCs) as well as an automated method for enumeration of RBCs and WBCs in body fluids. The purpose of the LH 700 Series is to separate the normal patient, with all normal system-generated parameters, from the patient who needs additional studies of any of these parameters. These studies might include further measurements of cell size and platelet distribution, manual WBC differential or any other definitive test that helps diagnose the patients condition. The purpose of the LH 500 Analyzer is to separate the normal patient, with all normal system-generated parameters, from the patient who needs additional studies. These studies include further measurements of cell size and cell distribution, or any other test that aids in diagnosing an abnormality

Recall: Z-3113-2017 · Initiated June 26, 2017

Recall

Recall Number
Z-3113-2017
Event Number
77830
Firm
Beckman Coulter Inc.
FEI Number
2050012
Product Code
GKZ
Status
Terminated
Root Cause
Labeling Change Control
Initiated
June 26, 2017
Terminated
September 14, 2023
Address
250 S Kraemer Blvd, Brea, CA, 92821-6232

Description

Coulter LH 750, Coulter LH 780, and Coulter LH 500 Analyzers Hematology Analyzers are quantitative, automated hematology analyzers and leukocyte differential counters For In Vitro Diagnostic used in clinical laboratories. They also provide automated Reticulocyte analysis and enumeration of nucleated red blood cells (NRBCs) as well as an automated method for enumeration of RBCs and WBCs in body fluids. The purpose of the LH 700 Series is to separate the normal patient, with all normal system-generated parameters, from the patient who needs additional studies of any of these parameters. These studies might include further measurements of cell size and platelet distribution, manual WBC differential or any other definitive test that helps diagnose the patients condition. The purpose of the LH 500 Analyzer is to separate the normal patient, with all normal system-generated parameters, from the patient who needs additional studies. These studies include further measurements of cell size and cell distribution, or any other test that aids in diagnosing an abnormality

Reason

Beckman Coulter has determined that additional clarification for the Blast Suspect messages is necessary. In rare situations, the LH 750, LH 780, LH 500 and HmX Analyzers may not flag or detect blasts in some blood samples.

Action

Beckman Coulter sent an Urgent Medical Device Recall letter dated June 30, 2017, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to share the information with their laboratory staff. If they forwarded any of the affected product to another laboratory, please provide them a copy of the letter. Customers with questions should contact the Customer Support Center at 800-528-7694.

Distribution

Worldwide Distribution - US (nationwide) Internationally to Albania, Algeria, Andorra, Angola, Antigua and Barbuda, Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belgium, Bermuda, Bosnia and Herzegovina, Brazil, Bulgaria, Burundi, Canada, China, Colombia, Congo, Croatia, Czech republic, Ecuador, Egypt, Ethiopia, Finland, France, French Polynesia, Germany, Greece, Guadeloupe, Guatemala, Honduras, Hong Kong, Hungary, Iceland, india, Indonesia, Italy, Japan, Iraq, Jordan. Kenya, Kuwait, Libya, Lithuania, Macau, Lebanon, Macedonia, Malawi, Malaysia, Mexico, Mongolia, Morocco, Mozambique, Netherlands, New Zealand, Nigeria, Pakistan, Panama, Paraguay, Philippines, Portugal, Puerto Rico, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sri Lanka, Swaziland, Sweden, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Turkey, UAE, UK, Uruguay, Venezuela, Vietnam, Virgin Islands, Yemen, Zambia, Zimbabwe.

Quantity

6763 units