FDA Recall Terminated

CIVCO Arm Support, REF 106015, distributed as: (a) Rigid Arm Support, Standard, Set Red (PN: 106015) (b) Rigid Adj. Arm Support, Set (PN: 106020) (c) PSR2-CR-DLR-2P Posirest-2 (PN: 106521) (d) PB2-BLR Posiboard-2 (PN: 109030) The Rigid Arm Supports are utilized on the Posibboard2 and Posirest assemblies, which are intended to position the patients upper extremities during breast and lung diagnostic and radiotherapy treatment..

Recall: Z-3082-2017 · Initiated August 10, 2017

Recall

Recall Number
Z-3082-2017
Event Number
77874
Firm
Med Tec Inc
FEI Number
1000143606
Product Code
KXH
Status
Terminated
Root Cause
Other
Initiated
August 10, 2017
Terminated
March 19, 2024
Address
1401 8th St SE PO Box 320, Orange City, IA, 51041-7463

Description

CIVCO Arm Support, REF 106015, distributed as: (a) Rigid Arm Support, Standard, Set Red (PN: 106015) (b) Rigid Adj. Arm Support, Set (PN: 106020) (c) PSR2-CR-DLR-2P Posirest-2 (PN: 106521) (d) PB2-BLR Posiboard-2 (PN: 109030) The Rigid Arm Supports are utilized on the Posibboard2 and Posirest assemblies, which are intended to position the patients upper extremities during breast and lung diagnostic and radiotherapy treatment..

Reason

Rigid Arm Supports (PN: 106015 and 106047), which were used as part of the Posiboard-2 and Posirest, and were distributed as PN: 106015, 106047, 106521, 106522, 109030, and 109040 between Ocotober 2009 and July 2012 may have a height difference of approximately 17mm.

Action

CIVCO sent an Urgent: Field Action Notice- Customer Communication letter dated July 20th, 2017. The field action notification letter was sent on August 10, 2017, by email requesting the consignee review their inventory and contact their sales representative for product replacement. Enclosed please find a notification letter which is required to be sent to all customers who received this product. Please contact your Insides Sales Representative concerning replacement instructions if you have any questions. You may reach your representative by calling CIVCO Radiotherapy, at 712-737-8688 between 8:00AM and 5:00PM CST. For further question, please call (319) 248-6537.

Distribution

Worldwide Distribution - US Distribution to the states of : MI, IL and TX., and to the countries of : Greece, Australia, Czech Republic, Hungary, Norway, Brazil, Italy, Switzerland ,UK, Austria, Slovenia, Croatia, Poland, Germany, Japan, France, Netherland, Sri Lanka, Pakistan, Bangladesh, and Israel

Quantity

101 units