Hamamatsu model L10951-35 Micro Focus x-ray generator. The x-ray generator is incorporated into a cabinet x-ray system that is used for a variety of studies.
Recall
- Recall Number
- Z-3076-2011
- Event Number
- 59760
- Firm
- Hamamatsu Corporation
- FEI Number
- 1000121818
- Product Code
- RBS
- Status
- Terminated
- Root Cause
- Radiation Control for Health and Safety Act
- Initiated
- July 7, 2011
- Posted
- September 13, 2011
- Terminated
- January 29, 2018
- Address
- 360 Foothill Rd, Bridgewater, NJ, 08807
Description
Hamamatsu model L10951-35 Micro Focus x-ray generator. The x-ray generator is incorporated into a cabinet x-ray system that is used for a variety of studies.
It was discovered about a potential issue of X-ray leakage with our Model # LI1 0951-35 Micro Focus X-Ray generator.
Hamamatsu Corporation sent a letter dated July 19, 2011, to the manufacturer and their subcontractor informing them of the affected product and what they propose to do. They will repair the units both in the field and those returned to our plant by installing additional lead shielding in the units. The lead shielding will be fixed to the unit by soldering the addition to the existing lead shield (see attachment 2). We expect to complete the repair of all units in the field by October 1, 2011. All new units to be shipped from our plant will be repaired before shipment. For further questions, please call (908) 231-0960 X 2300
Worldwide Distribution -- USA ( nationwide ) the countries of Aichi, India and Japan.
8