FDA Recall Terminated

VITEK 2 Compact 15, REF numbers: 27415 and 27415R software and reagent cards designed for the identification (ID) and antimicrobial susceptibility testing (AST) of bacteria and yeast.

Recall: Z-3075-2017 · Initiated July 12, 2017

Recall

Recall Number
Z-3075-2017
Event Number
77799
Firm
Biomerieux Inc
FEI Number
1950204
Product Code
LON
Status
Terminated
Root Cause
Other
Initiated
July 12, 2017
Terminated
December 16, 2020
Address
595 Anglum Rd, Hazelwood, MO, 63042-2320

Description

VITEK 2 Compact 15, REF numbers: 27415 and 27415R software and reagent cards designed for the identification (ID) and antimicrobial susceptibility testing (AST) of bacteria and yeast.

Reason

Customers have reported that some VITEK 2 cards are staying in preliminary status, not finalizing after ejection from the instrument, and not allowing cards in subsequent carousel slots to be processed. The issue was reported to occur on VITEK 2 Compact 15 and Compact 30 systems following a system software update to version 8.01.

Action

Consignees (Subsidiaries) of bioMrieux, impacted by the identified issue, were notified of the FSCA and corrective action (FSCA notification) electronically on 07/12/2017. The direct consignees (subsidiaries and distributors) have until 08/12/2017 to notify their respective impacted users. The directions to the user are as follows: Please implement the following actions at this time: " Confirm this letter has been distributed to, and reviewed by, all appropriate personnel within your organization. " Refrain from performing the VITEK 2 Systems Software 8.01 update. o A solution has been identified, and your local bioMrieux representative will contact you with further instruction in the coming weeks. " If you have updated VITEK 2 Systems Software to version 8.01 within the last seven (7) days, please contact your local bioMrieux representative. " Please store this letter with your bioMrieux instrument documentation. " Complete the Acknowledgement Form and return it to your local bioMrieux representative."

Distribution

Worldwide

Quantity

1866 units