Sysmex Stromatolyser-IM; a lytic reagent for use with the XE-2100, XE-2100L and XE-5000 Series Sysmex Automated Hematology Analyzers; the reagent solution is supplied in a 10-liter cubetainer (cardboard cube with a 10-liter plastic liner); Sysmex America, Inc., Mundelein, IL 60060; catalog number SIM-220A. The STROMATOLYSER-IM lysing reagent is used to destroy red blood cells to aid in the counting and characterization on immature white blood cells. This reagent is used on the Sysmex XE-Series and SE Series hematology analyzers.
Recall
- Recall Number
- Z-3063-2011
- Event Number
- 59523
- Firm
- Sysmex America, Inc.
- FEI Number
- 1000515253
- Product Code
- GGK
- Status
- Terminated
- Root Cause
- Process change control
- Initiated
- April 11, 2011
- Posted
- August 23, 2011
- Terminated
- March 14, 2012
- Address
- 1 Nelson C White Pkwy, Mundelein, IL, 60060-9528
Description
Sysmex Stromatolyser-IM; a lytic reagent for use with the XE-2100, XE-2100L and XE-5000 Series Sysmex Automated Hematology Analyzers; the reagent solution is supplied in a 10-liter cubetainer (cardboard cube with a 10-liter plastic liner); Sysmex America, Inc., Mundelein, IL 60060; catalog number SIM-220A. The STROMATOLYSER-IM lysing reagent is used to destroy red blood cells to aid in the counting and characterization on immature white blood cells. This reagent is used on the Sysmex XE-Series and SE Series hematology analyzers.
The affected lots of Stromatolyser-IM lysing reagent were found to be contaminated with Pseudomonas bacteria, which resulted in rising IMI channel background counts, increased false "Platelet Clump" flags and a strong sulfur-like odor when used on the XE-2100, XE-2100L and XE-5000 analyzers.
Sysmex America, Inc. sent a "PRODUCT NOTIFICATION" letter dated April 2011 to all affected customers. The letter identifies the product, problem, and actions to be taken. The letter instructs customers to discontinue use of the affected SIM lots, and to remove affected lots from their analyzers as soon as possible to avoid long-term performance problems on the analyzer. A Stromatolyser -IM (220-A) Voluntary Recall form was attached for customers to complete and return via fax to Sysmex Customer Service at 800-689-8296. The form will be used to place orders to replace the affected product. Any questions concerning the information contained in the notification are directed to the Sysmex Technical Assistance Center at 1-888-879-7639.
Worldwide Distribution--USA (nationwide and Puerto Rico) and the countries of Canada, Colombia, Ecuador, Mexico, Netherland Antilles and Peru.
1,887 cubetainers