FDA Recall Terminated

TruChoice Alliant Healthcare Products, REFERENCE# SSD-100EX, QUANTITY: 25, DESCRIPTION: TUBING W/ ONE WAY VALVE, STERILE The TrueChoice Vacuum Control One-Way Vent Valve is indicated for use during cardiopulmonary bypass surgery to help prevent the buildup of excess vacuum pressure when suctioning either from the heart or the surgical field, and to help prevent inadvertent retrograde flow and air to the heart.

Recall: Z-3062-2011 · Initiated July 21, 2011

Recall

Recall Number
Z-3062-2011
Event Number
59524
Firm
Alliant Enterprises LLC
FEI Number
3004986960
Product Code
MJJ
Status
Terminated
Root Cause
Other
Initiated
July 21, 2011
Posted
August 23, 2011
Terminated
April 23, 2012
Address
8850 M 89, Richland, MI, 49083-8558

Description

TruChoice Alliant Healthcare Products, REFERENCE# SSD-100EX, QUANTITY: 25, DESCRIPTION: TUBING W/ ONE WAY VALVE, STERILE The TrueChoice Vacuum Control One-Way Vent Valve is indicated for use during cardiopulmonary bypass surgery to help prevent the buildup of excess vacuum pressure when suctioning either from the heart or the surgical field, and to help prevent inadvertent retrograde flow and air to the heart.

Reason

The firm is initiating this recall due to the receipt of 2 complaints of malfunction of the joint holding the 2 halves of the vent valve together.

Action

Alliant Healthcare sent an URGENT: MEDICAL DEVICE RECALL letter on 7/21/2011 and 7/25/2011. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to immediately review the recall notification and quarantine any affected product. If product was further distributed, customers were to forward the letter with the Recall Return Response Form immediately to all affected locations and indicate which locations on the Recall Return Response Form. Customers were also instructed to discontinue use immediately, dispose of the affected product per their normal disposal protocol, and fax the Recall Return Response Form to Alliant Healthcare Products at 269-629-5770. Upon receipt of the Recall Return Response Form, a substitute vent valve (non-adjustable only) will be sent to their location for each affected product for which disposal has been properly documented in writing. For questions call 269-629-0300, ext 226.

Distribution

Worldwide Distribution - USA including Michigan, Tennessee, New York, and California and the countries of Chile and Korea

Quantity

16 Boxes of 25