FDA Recall Terminated

Various custom surgical kits manufactured for CMS, CUSTOM MEDICAL SPECIALTIES, INC.: (a) CMS-4153-R2 MPM Biopsy TRAY (b) CMS-5832-R1 Custom Safety Mammo Biopsy Tray (c) CMS-4108 Multipurpose Safety Tray (d) CMS-3141-R6 Custom Ultrasound Tray (e) CMS-2891-R1 Custom Safety Biopsy Tray (f) CMS-5798-R1 Custom Safety Biopsy Tray (g) CMS-2357R2 Custom Biopsy Tray (h) CMS-2540-R4 Custom Biopsy Tray

Recall: Z-3059-2017 · Initiated June 15, 2017

Recall

Recall Number
Z-3059-2017
Event Number
77962
Firm
Custom Assemblies Inc
FEI Number
3000718854
Product Code
LRO
Status
Terminated
Root Cause
Material/Component Contamination
Initiated
June 15, 2017
Posted
August 22, 2017
Terminated
August 28, 2017
Address
330 E. Main Street, Pine Level, NC, 27568

Description

Various custom surgical kits manufactured for CMS, CUSTOM MEDICAL SPECIALTIES, INC.: (a) CMS-4153-R2 MPM Biopsy TRAY (b) CMS-5832-R1 Custom Safety Mammo Biopsy Tray (c) CMS-4108 Multipurpose Safety Tray (d) CMS-3141-R6 Custom Ultrasound Tray (e) CMS-2891-R1 Custom Safety Biopsy Tray (f) CMS-5798-R1 Custom Safety Biopsy Tray (g) CMS-2357R2 Custom Biopsy Tray (h) CMS-2540-R4 Custom Biopsy Tray

Reason

The firm manufactured and distributed custom surgical kits which contained BD SafetyGlide" 22GA 1 inch Needles. BD subsequently recalled these needles because, " Loose polypropylene particulate in the fluid path could be expelled out of the needle during injection.

Action

BD sent an Urgent Medical Device Recall letter dated June 12, 2017, to all affected consignees. Consignees were instructed to examine their inventory and quarantine any affected product. Consignees with questions should call 201-847-4267. For questions regarding this recall call 919-202-8462.

Distribution

Nationwide Distribution to NJ, TN, NY, KS, OH, CA,

Quantity

2155 kits