The Resolution II Clip, 155 cm size and 235 cm size, Boxes of 1, 10, and 20 clips consists of a pre-loaded, radiopaque single use clip on a flexible delivery system. Endoscopic Clipping Device, Hemorrhoidal Ligator. The device is used for clip placement within the GI tract for hemostatis ( mucosal defects, bleeding ulcer, arteries, polyps); endoscopic marking; and anchoring to affix jejunal feeding tubes to the wall of the small bowel, as a supplementary method for closure of GI tract luminal perforation <20 mm that can be treated conservatively.

Recall: Z-3056-2011 · Initiated July 19, 2011

Recall

Recall Number
Z-3056-2011
Event Number
59476
Firm
Boston Scientific Corporation
FEI Number
3005099803
Product Code
FHN
Status
Terminated
Root Cause
Other
Initiated
July 19, 2011
Posted
August 22, 2011
Terminated
June 22, 2012
Address
100 Boston Scientific Way, Marlborough, MA, 01752-1234

Description

The Resolution II Clip, 155 cm size and 235 cm size, Boxes of 1, 10, and 20 clips consists of a pre-loaded, radiopaque single use clip on a flexible delivery system. Endoscopic Clipping Device, Hemorrhoidal Ligator. The device is used for clip placement within the GI tract for hemostatis ( mucosal defects, bleeding ulcer, arteries, polyps); endoscopic marking; and anchoring to affix jejunal feeding tubes to the wall of the small bowel, as a supplementary method for closure of GI tract luminal perforation <20 mm that can be treated conservatively.

Reason

Boston Scientific has received reports where, following clip deployment, the alignment of the clip and delivery system does not allow for the withdrawal of the delivery system away from the clip, resulting in tissue trauma, increased bleeding and the clip being pulled off tissue as the delivery system is withdrawn.

Action

Boston Scientific sent an URGENT MEDICAL DEVICE RECALL letter dated August 12, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to forward the Medical Device Recall to their customers. Customers were instructed to: Immediately discontinue use of any affected product. Complete and return the Reply Verification Tracking Form by fax to 508-683-5578. Package and return any affected product to: Boston Scientific Corporation Distribution Center Attn: QA Returns 500 Commander Shea Blvd Quincy, MA 02171 For questions regarding this recall call 508-683-5578.

Distribution

Worldwide Distribution - including USA (nationwide)

Quantity

14, 093 boxes of both 155 cm and 235 cm size clips