FDA Recall Terminated

Drug Screening Component: Uncut Sheets Part Number/ Description/ LOT Number 100163/ EDI OPI (300) Uncut Sheet-Card /187876, 100195/ EDI PPX Uncut Sheet-Card/ 187877, 160130/ EDI COC Uncut Sheet-Cup/ 188168, 160180/ EDI PCP Uncut Sheet-Cup/ 189027, 160140/ EDI THC Uncut Sheet Cup /189028, 160170/ EDI BZO Uncut Sheet-Cup/ 189081, 160120/ EDI MET1000 Uncut Sheet-Cup/ 189082, 160160/ EDI OPI2000 Uncut Sheet-Cup / 189123, 160130/ EDI COC Uncut Sheet-Cup/ 189124, 160160/ EDI OPI2000 Uncut Sheet-Cup/ 189166, 160163/ EDI OPI (300) Uncut Sheet-Cup/ 189191, 160110/ EDI AMP Uncut Sheet-Cup/ 189371, 160140/ EDI THC Uncut Sheet Cup/ 189372, 160123/ EDI MET(300) Uncut Sheet-Cup/ 189425, 160170/ EDI BZO Uncut Sheet-Cup/ 189426, 100163/ EDI OPI (300) Uncut Sheet-Card/ 189456, 160170/ EDI BZO Uncut Sheet-Cup/ 189565, 160180/ EDI PCP Uncut Sheet-Cup/ 189601, 160125/ EDI MET(500) Uncut Sheet-Cup/ 189748, 160130/ EDI COC Uncut Sheet-Cup/ 189786, 160160/ EDI OPI2000 Uncut Sheet-Cup / 189844, 160180/ EDI PCP Uncut Sheet-Cup/ 189937, 160172/ EDI BZO(200) Uncut Sheet-CUP/189956 Updated 09/03/2020: EDI Affected products -Catalog Numbers/ Lot Numbers 31105-3/ PI-01081 60503s-3ZC/ PH-10132 60600/ PH-11005

Recall: Z-3051-2020 · Initiated August 18, 2020

Recall

Recall Number
Z-3051-2020
Event Number
86335
Firm
Ameditech Inc
FEI Number
3003789989
Product Code
DIO
Status
Terminated
Root Cause
Process control
Initiated
August 18, 2020
Terminated
September 30, 2024
Address
9940 Mesa Rim Rd, San Diego, CA, 92121-2910

Description

Drug Screening Component: Uncut Sheets Part Number/ Description/ LOT Number 100163/ EDI OPI (300) Uncut Sheet-Card /187876, 100195/ EDI PPX Uncut Sheet-Card/ 187877, 160130/ EDI COC Uncut Sheet-Cup/ 188168, 160180/ EDI PCP Uncut Sheet-Cup/ 189027, 160140/ EDI THC Uncut Sheet Cup /189028, 160170/ EDI BZO Uncut Sheet-Cup/ 189081, 160120/ EDI MET1000 Uncut Sheet-Cup/ 189082, 160160/ EDI OPI2000 Uncut Sheet-Cup / 189123, 160130/ EDI COC Uncut Sheet-Cup/ 189124, 160160/ EDI OPI2000 Uncut Sheet-Cup/ 189166, 160163/ EDI OPI (300) Uncut Sheet-Cup/ 189191, 160110/ EDI AMP Uncut Sheet-Cup/ 189371, 160140/ EDI THC Uncut Sheet Cup/ 189372, 160123/ EDI MET(300) Uncut Sheet-Cup/ 189425, 160170/ EDI BZO Uncut Sheet-Cup/ 189426, 100163/ EDI OPI (300) Uncut Sheet-Card/ 189456, 160170/ EDI BZO Uncut Sheet-Cup/ 189565, 160180/ EDI PCP Uncut Sheet-Cup/ 189601, 160125/ EDI MET(500) Uncut Sheet-Cup/ 189748, 160130/ EDI COC Uncut Sheet-Cup/ 189786, 160160/ EDI OPI2000 Uncut Sheet-Cup / 189844, 160180/ EDI PCP Uncut Sheet-Cup/ 189937, 160172/ EDI BZO(200) Uncut Sheet-CUP/189956 Updated 09/03/2020: EDI Affected products -Catalog Numbers/ Lot Numbers 31105-3/ PI-01081 60503s-3ZC/ PH-10132 60600/ PH-11005

Reason

Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.

Action

A Field Corrective Action Notice letter was sent to affected consignees. The Customer notifications commenced on 19-Aug-2020. The consignees are requested to return all non-expired product in their possession to the recalling firm. Method of Notification, 1st contact: Written field corrective action notification letters will be sent via certified mail with delivery tracking to each consignee along with a pre-paid return label to return any product remaining in stock. 2nd & 3rd contact: Telephone, follow up email, or written mail communication will be used as necessary.

Distribution

US - Nationwide OUS - Australia, France, Germany, New Zealand, Singapore and Spain

Quantity

3600 units