FDA Recall Open, Classified

Medline Convenience kits labeled as: 1) PACK VAGINAL DELIVERY , Pack Number AM185B ; 2) LABOR & DELIVERY CDS, Pack Number CDS830039AF; 3) LABOR & DELIVERY CDS, Pack Number CDS830039AI; 4) LABOR & DELIVERY CDS, Pack Number CDS830039AJ; 5) LABOR & DELIVERY CDS, Pack Number CDS830039AK; 6) VAGINAL DELIVERY CDS, Pack Number CDS830053G ; 7) VAGINAL DELIVERY CDS, Pack Number CDS830053I ; 8) LABOR & DELIVERY, Pack Number CDS830054P ; 9) VAGINAL DELIVERY PACK-LF, Pack Number CDS830116T ; 10) VAG DELIVERY, ROSCOE, Pack Number CDS830118Q ; 11) VAG DELIVERY, ROSCOE, Pack Number CDS830118R ; 12) VAG DELIVERY, ROSCOE, Pack Number CDS830118S ; 13) VAG DELIVERY, ROSCOE, Pack Number CDS830118T ; 14) VAGINAL DEL/RECOV CDS , Pack Number CDS830131J ; 15) VAGINAL DEL/RECOV CDS , Pack Number CDS830131K ; 16) VAGINAL DEL/RECOV CDS , Pack Number CDS830131L ; 17) MAJOR VAGINAL PROCEDURE , Pack Number CDS830136S ; 18) VAGINAL DELIVERY PROCEDURE, Pack Number CDS830137O ; 19) VAGINAL DELIVERY PROCEDURE, Pack Number CDS830137P ; 20) VAGINAL DELIVERY CDS, Pack Number CDS830140A ; 21) VAGINAL DELIVERY CDS, Pack Number CDS830140B ; 22) VAGINAL DELIVERY CDS, Pack Number CDS830178K ; 23) VAGINAL DELIVERY CDS, Pack Number CDS830178L ; 24) VAGINAL DELIVERY CDS, Pack Number CDS830178M ; 25) VAGINAL HYSTERECTOMY CDS, Pack Number CDS830186O ; 26) DELIVERY CDS, Pack Number CDS830199K ; 27) DELIVERY CDS, Pack Number CDS981282O ; 28) OB PROCEDURE-LF , Pack Number CDS981335B ; 29) VAG HYSTERECTOMY CDS, Pack Number CDS981517G ; 30) VAGINAL DELIVERY CDS, Pack Number CDS981776I ; 31) VAGINAL DELIVERY CDS, Pack Number CDS981776J ; 32) VAGINAL DELIVERY CDS, Pack Number CDS981776K ; 33) VAGINAL HYSTERECTOMY CDS, Pack Number CDS982369F ; 34) VAGINAL MAJOR , Pack Number CDS982689O ; 35) VAGINAL MAJOR , Pack Number CDS982689P ; 36) VAGINAL MAJOR , Pack Number CDS982689Q ; 37) VAGINAL MAJOR , Pack Number CDS982689R ; 38) DELIVERY PROCEDURE, Pack Number CDS982761F ; 39) DELIVERY PROCEDURE, Pack Number CDS982761G ; 40) DELIVERY PROCEDURE, Pack Number CDS982761I ; 41) DELIVERY PROCEDURE, Pack Number CDS982761J ; 42) VAGINAL DELIVERY CDS, Pack Number CDS982762D ; 43) VAGINAL DELIVERY CDS, Pack Number CDS982762F ; 44) VAGINAL DELIVERY CDS, Pack Number CDS982762G ; 45) VAGINAL DELIVERY CDS, Pack Number CDS982762I ; 46) VAGINAL DELIVERY CDS, Pack Number CDS982770B ; 47) MAG MAJOR VAGINAL CDS , Pack Number CDS982907K ; 48) MAG MAJOR VAGINAL CDS , Pack Number CDS982907L ; 49) MAG MINOR VAGINAL CDS , Pack Number CDS982908I ; 50) MAG MINOR VAGINAL CDS , Pack Number CDS982908J ; 51) OB-VAGINAL DELIVERY/EMERGENCY , Pack Number CDS982959; 52) VAGINAL DELIVERY CDS, Pack Number CDS983090C ; 53) VAGINAL HYSTERECTOMY CDS, Pack Number CDS983107G ; 54) LABOR & DELIVERY CDS, Pack Number CDS983116C ; 55) LABOR & DELIVERY CDS, Pack Number CDS983116D ; 56) LABOR & DELIVERY CDS, Pack Number CDS983116F ; 57) LABOR & DELIVERY CDS, Pack Number CDS983116G ; 58) LABOR & DELIVERY CDS, Pack Number CDS983116I ; 59) MINOR VAGINAL CDS , Pack Number CDS983119B ; 60) VAGINAL MINOR , Pack Number CDS983268J ; 61) VAGINAL MINOR , Pack Number CDS983268L ; 62) GYN VAGINAL CDS , Pack Number CDS983528J ; 63) ACH MAJOR VAGINAL , Pack Number CDS983641F ; 64) ACH MAJOR VAGINAL , Pack Number CDS983641G ; 65) VAG HYST CDS, Pack Number CDS983697F ; 66) VAGINAL DELIVERY CDS, Pack Number CDS983919C ; 67) VAGINAL DELIVERY CDS, Pack Number CDS983919D ; 68) VAGINAL DELIVERY CDS, Pack Number CDS983919F ; 69) VAGINAL DELIVERY CDS, Pack Number CDS983919G ; 70) OB/GYN CDS, Pack Number CDS984173G ; 71) OB GYN, Pack Number CDS984173I ; 72) GROTH VAGINAL CDS , Pack Number CDS984201P ; 73) GROTH VAGINAL CDS , Pack Number CDS984201Q ; 74) GROTH VAGINAL CDS , Pack Number CDS984201R ; 75) GROTH VAGINAL CDS , Pack Number CDS984201U ; 76) NHP VAGINAL MINOR CDS , Pack Number CDS984235F ; 77) VAGINAL DELIVERY CDS, Pack Number CDS984352C ; 78) VAGINAL DELIVERY CDS, Pack

Recall: Z-3044-2024 · Initiated April 8, 2024

Recall

Recall Number
Z-3044-2024
Event Number
94583
Firm
MEDLINE INDUSTRIES, LP Northfield
FEI Number
1417592
Product Code
MLS
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
April 8, 2024
Posted
September 6, 2024
Address
3 Lakes Dr, Northfield, IL, 60093-2753

Description

Medline Convenience kits labeled as: 1) PACK VAGINAL DELIVERY , Pack Number AM185B ; 2) LABOR & DELIVERY CDS, Pack Number CDS830039AF; 3) LABOR & DELIVERY CDS, Pack Number CDS830039AI; 4) LABOR & DELIVERY CDS, Pack Number CDS830039AJ; 5) LABOR & DELIVERY CDS, Pack Number CDS830039AK; 6) VAGINAL DELIVERY CDS, Pack Number CDS830053G ; 7) VAGINAL DELIVERY CDS, Pack Number CDS830053I ; 8) LABOR & DELIVERY, Pack Number CDS830054P ; 9) VAGINAL DELIVERY PACK-LF, Pack Number CDS830116T ; 10) VAG DELIVERY, ROSCOE, Pack Number CDS830118Q ; 11) VAG DELIVERY, ROSCOE, Pack Number CDS830118R ; 12) VAG DELIVERY, ROSCOE, Pack Number CDS830118S ; 13) VAG DELIVERY, ROSCOE, Pack Number CDS830118T ; 14) VAGINAL DEL/RECOV CDS , Pack Number CDS830131J ; 15) VAGINAL DEL/RECOV CDS , Pack Number CDS830131K ; 16) VAGINAL DEL/RECOV CDS , Pack Number CDS830131L ; 17) MAJOR VAGINAL PROCEDURE , Pack Number CDS830136S ; 18) VAGINAL DELIVERY PROCEDURE, Pack Number CDS830137O ; 19) VAGINAL DELIVERY PROCEDURE, Pack Number CDS830137P ; 20) VAGINAL DELIVERY CDS, Pack Number CDS830140A ; 21) VAGINAL DELIVERY CDS, Pack Number CDS830140B ; 22) VAGINAL DELIVERY CDS, Pack Number CDS830178K ; 23) VAGINAL DELIVERY CDS, Pack Number CDS830178L ; 24) VAGINAL DELIVERY CDS, Pack Number CDS830178M ; 25) VAGINAL HYSTERECTOMY CDS, Pack Number CDS830186O ; 26) DELIVERY CDS, Pack Number CDS830199K ; 27) DELIVERY CDS, Pack Number CDS981282O ; 28) OB PROCEDURE-LF , Pack Number CDS981335B ; 29) VAG HYSTERECTOMY CDS, Pack Number CDS981517G ; 30) VAGINAL DELIVERY CDS, Pack Number CDS981776I ; 31) VAGINAL DELIVERY CDS, Pack Number CDS981776J ; 32) VAGINAL DELIVERY CDS, Pack Number CDS981776K ; 33) VAGINAL HYSTERECTOMY CDS, Pack Number CDS982369F ; 34) VAGINAL MAJOR , Pack Number CDS982689O ; 35) VAGINAL MAJOR , Pack Number CDS982689P ; 36) VAGINAL MAJOR , Pack Number CDS982689Q ; 37) VAGINAL MAJOR , Pack Number CDS982689R ; 38) DELIVERY PROCEDURE, Pack Number CDS982761F ; 39) DELIVERY PROCEDURE, Pack Number CDS982761G ; 40) DELIVERY PROCEDURE, Pack Number CDS982761I ; 41) DELIVERY PROCEDURE, Pack Number CDS982761J ; 42) VAGINAL DELIVERY CDS, Pack Number CDS982762D ; 43) VAGINAL DELIVERY CDS, Pack Number CDS982762F ; 44) VAGINAL DELIVERY CDS, Pack Number CDS982762G ; 45) VAGINAL DELIVERY CDS, Pack Number CDS982762I ; 46) VAGINAL DELIVERY CDS, Pack Number CDS982770B ; 47) MAG MAJOR VAGINAL CDS , Pack Number CDS982907K ; 48) MAG MAJOR VAGINAL CDS , Pack Number CDS982907L ; 49) MAG MINOR VAGINAL CDS , Pack Number CDS982908I ; 50) MAG MINOR VAGINAL CDS , Pack Number CDS982908J ; 51) OB-VAGINAL DELIVERY/EMERGENCY , Pack Number CDS982959; 52) VAGINAL DELIVERY CDS, Pack Number CDS983090C ; 53) VAGINAL HYSTERECTOMY CDS, Pack Number CDS983107G ; 54) LABOR & DELIVERY CDS, Pack Number CDS983116C ; 55) LABOR & DELIVERY CDS, Pack Number CDS983116D ; 56) LABOR & DELIVERY CDS, Pack Number CDS983116F ; 57) LABOR & DELIVERY CDS, Pack Number CDS983116G ; 58) LABOR & DELIVERY CDS, Pack Number CDS983116I ; 59) MINOR VAGINAL CDS , Pack Number CDS983119B ; 60) VAGINAL MINOR , Pack Number CDS983268J ; 61) VAGINAL MINOR , Pack Number CDS983268L ; 62) GYN VAGINAL CDS , Pack Number CDS983528J ; 63) ACH MAJOR VAGINAL , Pack Number CDS983641F ; 64) ACH MAJOR VAGINAL , Pack Number CDS983641G ; 65) VAG HYST CDS, Pack Number CDS983697F ; 66) VAGINAL DELIVERY CDS, Pack Number CDS983919C ; 67) VAGINAL DELIVERY CDS, Pack Number CDS983919D ; 68) VAGINAL DELIVERY CDS, Pack Number CDS983919F ; 69) VAGINAL DELIVERY CDS, Pack Number CDS983919G ; 70) OB/GYN CDS, Pack Number CDS984173G ; 71) OB GYN, Pack Number CDS984173I ; 72) GROTH VAGINAL CDS , Pack Number CDS984201P ; 73) GROTH VAGINAL CDS , Pack Number CDS984201Q ; 74) GROTH VAGINAL CDS , Pack Number CDS984201R ; 75) GROTH VAGINAL CDS , Pack Number CDS984201U ; 76) NHP VAGINAL MINOR CDS , Pack Number CDS984235F ; 77) VAGINAL DELIVERY CDS, Pack Number CDS984352C ; 78) VAGINAL DELIVERY CDS, Pack

Reason

Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.

Action

Medline Industries LP issued a recall notice to its consignees on 04/08/2024 via letter delivered via USPS and email. The notice explained the issue, provided a link to for information pertaining to the risk involved https://www.fda.gov/medical-devices/safety-communications/update-evaluating-plastic-syringes-made-china-potential-device-failures-fda-safety-communication) and request these syringes within the convenience be discarded, except for circumstances where no alternative syringes are readily available. If these syringes must be used, closely monitor for leaks, breakage, and any other problems, and report any quality issues to Medline and the FDA. Distributors were directed to notify their customers. The firm issued a second recall notice (an expansion) on 05/02/2024 using the same methods.

Distribution

Worldwide distribution.