FDA Recall Terminated

HomeChoice Pro Automated PD system Model numbers: 5C8310, 5C8310R Product Usage: The HomeChoice Automated Personal Cycler is intended for automatic control of dialysate solution exchanges in the treatment of pediatric and adult renal failure patients undergoing peritoneal dialysis.

Recall: Z-3041-2017 · Initiated July 5, 2017

Recall

Recall Number
Z-3041-2017
Event Number
77711
Firm
Baxter Healthcare Corporation
FEI Number
1417572
Product Code
FKX
Status
Terminated
Root Cause
Use error
Initiated
July 5, 2017
Terminated
October 5, 2020
Address
1 Baxter Pkwy, Deerfield, IL, 60015-4625

Description

HomeChoice Pro Automated PD system Model numbers: 5C8310, 5C8310R Product Usage: The HomeChoice Automated Personal Cycler is intended for automatic control of dialysate solution exchanges in the treatment of pediatric and adult renal failure patients undergoing peritoneal dialysis.

Reason

Baxter Healthcare Corporation has been made aware that users may not be following the instructions in the Operators Manual and incorrectly (using sharp objects) opening disposable set packaging while setting up their Peritoneal Dialysis (PD) therapy, damaging the cassettes for the HomeChoice or HomeChoice PRO cyclers. If damaged cassettes are used, the cyclers may not consistently detect very small holes/cuts in the sheeting of the cassette in the patient valve region and the cycler may deliver air into the patient.

Action

The firm notified their consignees by letter on 07/05/2017 of the potential for damaging the disposable set if tools are used to open the packaging and directing them to only open the packaging by hand by letter.

Distribution

Worldwide Distribution

Quantity

23732 units