FDA Recall Terminated

EMBLEM MRI S-ICD Model 219, Subcutaneous-Implantable Cardioverter Defibrillator, Rx. Product Usage: The S-ICD system is intended to provide defibrillation therapy for the treatment of life-threatening ventricular tachyarrhythmias in patients who do not have symptomatic bradycardia, incessant ventricular tachycardia, or spontaneous, frequently recurring ventricular tachycardia that is reliably terminated with anti-tachycardia pacing

Recall: Z-3038-2017 · Initiated June 29, 2017

Recall

Recall Number
Z-3038-2017
Event Number
77803
Firm
Boston Scientific Corporation
FEI Number
2124215
Product Code
LWS
Status
Terminated
Root Cause
Software design
Initiated
June 29, 2017
Posted
August 16, 2017
Terminated
February 21, 2019
Address
4100 Hamline Ave N, Saint Paul, MN, 55112-5700

Description

EMBLEM MRI S-ICD Model 219, Subcutaneous-Implantable Cardioverter Defibrillator, Rx. Product Usage: The S-ICD system is intended to provide defibrillation therapy for the treatment of life-threatening ventricular tachyarrhythmias in patients who do not have symptomatic bradycardia, incessant ventricular tachycardia, or spontaneous, frequently recurring ventricular tachycardia that is reliably terminated with anti-tachycardia pacing

Reason

The device can deliver an atypical amount of energy due to memory corruption inside the device.

Action

The firm issued notifications dated June, 2017, beginning 6/29/2017 to physicians (implanting and patient follow-up physicians) via hand delivery by their sales representatives. The firm estimated approximately 30% of the U.S. physicians would be receiving the notification via hand delivery. Hand delivery by affiliates in foreign countries to customers began approximately 6/30/2017. Overnight mail letters were issued to the U.S. and foreign physicians starting on/about 7/7/2017 who did not receive hand- delivered notifications.

Distribution

Worldwide - US Nationwide distribution, including Puerto Rico, U.S. Virgin Island, and Guam, There was also worldwide foreign distribution, including Canada.

Quantity

Approximately 9,200 devices