Henry Schein One Step + DOA Cup (AMT Item Number/ AMT Description): 570-1066 / "Henry Schein OneStep+ DOA Cup 14 Panel AMP500/BAR/BUPG/BZO/COC150/MDMA/MET500/MOP/MTD/OXY/PCP/PPX/TCA/THC (Bl, Cr, Ni, pH, SG)",
Recall
- Recall Number
- Z-3035-2020
- Event Number
- 86335
- Firm
- Ameditech Inc
- FEI Number
- 3003789989
- Product Code
- LDJ
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- August 18, 2020
- Terminated
- September 30, 2024
- Address
- 9940 Mesa Rim Rd, San Diego, CA, 92121-2910
Description
Henry Schein One Step + DOA Cup (AMT Item Number/ AMT Description): 570-1066 / "Henry Schein OneStep+ DOA Cup 14 Panel AMP500/BAR/BUPG/BZO/COC150/MDMA/MET500/MOP/MTD/OXY/PCP/PPX/TCA/THC (Bl, Cr, Ni, pH, SG)",
Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.
A Field Corrective Action Notice letter was sent to affected consignees. The Customer notifications commenced on 19-Aug-2020. The consignees are requested to return all non-expired product in their possession to the recalling firm. Method of Notification, 1st contact: Written field corrective action notification letters will be sent via certified mail with delivery tracking to each consignee along with a pre-paid return label to return any product remaining in stock. 2nd & 3rd contact: Telephone, follow up email, or written mail communication will be used as necessary.
US - Nationwide OUS - Australia, France, Germany, New Zealand, Singapore and Spain
31,800 units