PERPOS PLS, CATALOG #'s 9045-01 & 9045-02 A sterile, medical device kit consisting of the components listed in Table 1-1 below. The kit is used by medical professionals to stabilize the spine as an aid to fusion through immobilization of the facet joints.
Recall
- Recall Number
- Z-3033-2011
- Event Number
- 58381
- Firm
- Interventional Spine Inc
- FEI Number
- 3003523239
- Product Code
- MRW
- Status
- Terminated
- Root Cause
- Other
- Initiated
- March 15, 2011
- Posted
- August 18, 2011
- Terminated
- February 11, 2013
- Address
- 13700 Alton Pkwy, Ste 160, Irvine, CA, 92618-1618
Description
PERPOS PLS, CATALOG #'s 9045-01 & 9045-02 A sterile, medical device kit consisting of the components listed in Table 1-1 below. The kit is used by medical professionals to stabilize the spine as an aid to fusion through immobilization of the facet joints.
This recall has been initiated due to Pull Pins that may disengage from the Facet Screw during the compression step, requiring that compression be completed with a device driver rather than the compression tool. Use of this product may result in less than optimal compression leading to pseudoarthrosis, non-union or facet pain.
Interventional Spine, Inc. sent a recall letter dated March 15, 2011, informing consignees of the product, the problem, and the action to be taken. Consignees were asked to immediately examine their inventory and quarantine product subject to recall. In addition, if they distributed the product further they were asked to notify Jane Metcalf, the Vice President of Quality Assurance at once. Interventional Spine further informed consignees that the recall was being carried out to the individual consignee level. And their assistance was appreciated and necessary to prevent use of this product. They were then asked to complete and return the enclosed response form as soon as possible. Questions were directed to call their regional directors. For questions regarding this recall call 949-472-0006.
Nationwide Distribution including CO, CT, FL, KY, LA, MA, MO, NC, NY, OK, OR, TN, TX, VA & WI
553 units