FDA Recall Terminated

Hemoglobin F & A2 Calibrator, 0.2 mL, 4 brown glass vials/box. Mfr. By: Tosoh Bioscience, Minato-ku, Tokyo, Japan. Hemoglobin F & A2 Calibrator is used as in-vitro calibration material for Beta-thalassemia testing on the Tosoh G7.

Recall: Z-3030-2011 · Initiated May 25, 2011

Recall

Recall Number
Z-3030-2011
Event Number
59488
Firm
Tosoh Bioscience Inc
FEI Number
3005529799
Product Code
JPD
Status
Terminated
Root Cause
Process control
Initiated
May 25, 2011
Posted
August 17, 2011
Terminated
February 24, 2012
Address
3600 Gantz Road Grove, City, OH, 43123

Description

Hemoglobin F & A2 Calibrator, 0.2 mL, 4 brown glass vials/box. Mfr. By: Tosoh Bioscience, Minato-ku, Tokyo, Japan. Hemoglobin F & A2 Calibrator is used as in-vitro calibration material for Beta-thalassemia testing on the Tosoh G7.

Reason

The value after reconstitution appeared to be lower than the assigned value on the label. The measured value for Lot BCAL05R was 5.3%, whereas the value on the label is 5.7%. The criteria of the value is +/- 0.3%.

Action

TOSOH Bioscience sent an "Urgent Recall Notification" letter dated July, 25, 2011 to all affected customers. The letter described the problem and asks customers to immediately discontinue use of the product and return all remaining product back to the firm. Customers can contact TOSOH for replacement. For additional information or question on this recall please contact TOSOH Technical Support at (800) 248-6769.

Distribution

Worldwide Distribution: Nationwide (USA) distribution including the states of FL and IL; and the country of Canada.

Quantity

6 boxes/4/2 mL btls