FDA Recall Terminated

UIBC; Catalog Number: OSR61205; UDI (01) 15099590011925; IVD; Rx Only; WARNING H316 H317 H351 H373 H411 - Product Usage: System reagent for the quantitative determination of Unsaturated Iron Binding Capacity (UIBC) in human serum and plasma on Beckman Coulter AU analyzers.

Recall: Z-3028-2020 · Initiated August 18, 2020

Recall

Recall Number
Z-3028-2020
Event Number
86273
Firm
Beckman Coulter Inc.
FEI Number
2050012
Product Code
JMO
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
August 18, 2020
Terminated
September 8, 2021
Address
250 S Kraemer Blvd, Brea, CA, 92821-6232

Description

UIBC; Catalog Number: OSR61205; UDI (01) 15099590011925; IVD; Rx Only; WARNING H316 H317 H351 H373 H411 - Product Usage: System reagent for the quantitative determination of Unsaturated Iron Binding Capacity (UIBC) in human serum and plasma on Beckman Coulter AU analyzers.

Reason

The manufacturer has identified that the use of Unsaturated Iron Binding Capacity (UBIC) lot 2690 will produce erroneously high patient results at the low end of the linear range.

Action

An Urgent Medical Device Recall letter was sent on 08/18/2020 to the affected customers via mail and/or email. The customers are advised to discontinue the use of UIBC lot 2690. When available, replace it with an unaffected lot of UIBC.

Distribution

Worldwide distribution - US Nationwide distribution including in the states of CA, FL and NJ. The countries of Algeria, Egypt, Philippines and Taiwan.

Quantity

219 units