FDA Recall Terminated

1.5T Signa HDxt, 3.0T Signa HDxt, Signa HDi, Signa Vibrant Product Usage: The 1.5T Signa HDx family and 3.0T Signa HDx family are a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. Depending on the region of interest being imaged, contrast agents may be used. The images produced by the 1.5T Signa HDx family and 3.0T Signa HDx family reflects the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.

Recall: Z-3028-2017 · Initiated May 11, 2017

Recall

Recall Number
Z-3028-2017
Event Number
77488
Firm
GE Medical Systems, LLC
FEI Number
2183553
Product Code
LNH
Status
Terminated
Root Cause
Mixed-up of materials/components
Initiated
May 11, 2017
Posted
August 2, 2017
Terminated
September 9, 2020
Address
3200 N Grandview Blvd, Waukesha, WI, 53188-1693

Description

1.5T Signa HDxt, 3.0T Signa HDxt, Signa HDi, Signa Vibrant Product Usage: The 1.5T Signa HDx family and 3.0T Signa HDx family are a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. Depending on the region of interest being imaged, contrast agents may be used. The images produced by the 1.5T Signa HDx family and 3.0T Signa HDx family reflects the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.

Reason

Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.

Action

Consignees received a written recall notice dated 05/17/2017 notifying them of a correction to all units using an Uninterruptable Power Supply. Consignees were instructed not to unplug the unit and to continue normal use. Units with the Eaton 9130 Uninterruptable Power Supply connected to the MR operator console would receive a correction at no charge. This would involve possibly replacing the power cable. GE Representatives will be contacting consignees to arrange corrections. Any questions can be directed to GE Healthcare Service at 1-800-437-1171 or your local service representative.

Distribution

US Nationwide in the states of : AK, CA, CO, CT, DE, FL, IA, IL, IN, KS, KY, LA, MD, MO, ND, NE, NJ, NY, OH, PA, TX, WA, and WI

Quantity

68 units in total