BD Vacutainer Push Button Blood Collection Set; REF 367324 Rx only, 23 G x 3/4" x 12 " (0.6 x 19 mm x 305 mm); BD Vacutainer Push Button Blood Collection Set without Multiple Sample Luer Adaptor, Sterile, Rx only, 21 G x 3/4 x 12" 0.8 x 19 mm x 305 mm Blood collection system.
Recall
- Recall Number
- Z-3028-2011
- Event Number
- 59218
- Firm
- Becton Dickinson & Company
- FEI Number
- 2243072
- Product Code
- KST
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- April 22, 2011
- Posted
- August 16, 2011
- Terminated
- June 6, 2012
- Address
- 1 Becton Dr, Franklin Lakes, NJ, 07417-1815
Description
BD Vacutainer Push Button Blood Collection Set; REF 367324 Rx only, 23 G x 3/4" x 12 " (0.6 x 19 mm x 305 mm); BD Vacutainer Push Button Blood Collection Set without Multiple Sample Luer Adaptor, Sterile, Rx only, 21 G x 3/4 x 12" 0.8 x 19 mm x 305 mm Blood collection system.
BD Vacutainer Push Button Blood Collection Set can potentially undergo cannula separation from the hub.
BD Diagnostics sent a "Lot Specific Product Recall" letter dated April 22, 2011 to all affected customers. The letter asked customers return any affected product and provides instructions for its return. The letter ask customers return the attached Packaging List with the shipment. For question on this recall contact BD Diagnostics at (201) 847-5117.
Nationwide (USA) Distribution.
1,744,000