FDA Recall Terminated

Spacelabs Medical elance Vital Signs Monitor, Model 93300. The Spacelabs elance Vital Signs Monitor and elance Central Station is a prescription device intended for use by healthcare professionals in all areas of a healthcare facility. It is indicated for use in adult and pediatric populations. The device facilitates the monitoring of: ECG with arrhythmia detection; respiration; non-invasive blood pressures; invasive blood pressures; body temperature; functional arterial oxygen saturation; and entidal CO2.

Recall: Z-3024-2011 · Initiated September 23, 2010

Recall

Recall Number
Z-3024-2011
Event Number
56839
Firm
Spacelabs Healthcare, Llc
FEI Number
3010157426
Product Code
MHX
Status
Terminated
Root Cause
Error in labeling
Initiated
September 23, 2010
Posted
August 15, 2011
Terminated
August 17, 2011
Address
5150 220th Ave Se, Issaquah, WA, 98029-6834

Description

Spacelabs Medical elance Vital Signs Monitor, Model 93300. The Spacelabs elance Vital Signs Monitor and elance Central Station is a prescription device intended for use by healthcare professionals in all areas of a healthcare facility. It is indicated for use in adult and pediatric populations. The device facilitates the monitoring of: ECG with arrhythmia detection; respiration; non-invasive blood pressures; invasive blood pressures; body temperature; functional arterial oxygen saturation; and entidal CO2.

Reason

The monitor has a 3 second delay in the waveforms and respiration alarm; the delay is not disclosed in the IFU.

Action

Spacelabs Healthcare sent an "URGENT-MEDICAL DEVICE CORRECTION" letter dated September 24, 2010 to all affected customers. The letter describes the product, problem, and the action to be taken. The letter includes the revised Operator's Manual as a replacement for customers current manual. For Technical Support call Spacelabs Medical at 1-800-522-7025 and select 2.

Distribution

Worldwide Distribution-USA (Puerto Rico) and countries of ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BANGLADESH, BELGIUM, BOLIVIA, CHILE, CHINA, COLOMBIA, CYPRUS, CZECH REPUBLIC, DOMINICAN REPUBLIC, ECUADOR, EGYPT, EL SALVADOR, FRANCE, GERMANY, GREECE, GUATEMALA, HONDURAS, HONG KONG, INDIA, INDONESIA, IRAQ, ISRAEL, ITALY, KENYA, KOREA, REPUBLIC OF, KUWAIT, LEBANON, MACEDONIA, MALAYSIA, NEPAL, NETHERLANDS, NICARAGUA, OMAN, PALESTINE, PANAMA, PARAGUAY, PERU, PUERTO RICO, PHILIPPINES, QATAR, ROMANIA, SAUDI ARABIA, SINGAPORE, SOUTH AFRICA, SPAIN, SRI LANKA, SWITZERLAND, SYRIAN ARAB REPUBLIC, THAILAND, TRINIDAD AND TOBAGO, TUNISIA, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, VENEZUELA, and VIET NAM.

Quantity

2517