FDA Recall Terminated

fMRI Hardware System VisualSystem allows video signals from the stimulus presentation PC to enter the shielded scanner room and presented by a set of coil-mounted displays or by an in-room LCD monitor. - LCD Monitor Power supply (LPS) for LCD Monitor, Versions 1.0 and 1.1

Recall: Z-3023-2018 · Initiated August 20, 2018

Recall

Recall Number
Z-3023-2018
Event Number
80529
Firm
NordicNeuroLab AS Mollendalsveien 65c Bergen Norway
FEI Number
3006738448
Product Code
LNH
Status
Terminated
Root Cause
Labeling design
Initiated
August 20, 2018
Terminated
September 16, 2020

Description

fMRI Hardware System VisualSystem allows video signals from the stimulus presentation PC to enter the shielded scanner room and presented by a set of coil-mounted displays or by an in-room LCD monitor. - LCD Monitor Power supply (LPS) for LCD Monitor, Versions 1.0 and 1.1

Reason

Complaints of this version of the power supply unit (LPS) (1.0 and 1.1) being dragged into the magnet bore (MR Scanner).

Action

The firm sent a letter explaining the issue by email to the consignees on 08/20/2018. The firm requested that the consignee check the stock without removing the device from the magnet. It is was not in the magnet room, safely quarantine the device, and respond to the recall notice. The firm also directed any consignee who further distributed the product to notify those entities and to contact NNL. NNL will follow-up with them directly. NNL will upgrade all existing models with a safer design of the LCD Power supply unit.

Distribution

worldwide

Quantity

163 units