FDA Recall Terminated

Animas 2020 Insulin Infusion Pump. Animas Corp. An insulin infusion.

Recall: Z-3017-2011 · Initiated July 29, 2011

Recall

Recall Number
Z-3017-2011
Event Number
59517
Firm
Animas Corporation
FEI Number
3002920072
Product Code
LZG
Status
Terminated
Root Cause
Mixed-up of materials/components
Initiated
July 29, 2011
Posted
August 12, 2011
Terminated
December 7, 2011
Address
200 Lawrence Dr, West Chester, PA, 19380-3428

Description

Animas 2020 Insulin Infusion Pump. Animas Corp. An insulin infusion.

Reason

Incorrect component (1nF capacitor) was used to produce 191 printed circuit boards used in six pump lots of which 380 pumps were distributed. This issue may result in the pump displaying a "Replace Battery" alarm not related to the condition of the battery.

Action

Animas Corporation sent an "Urgent: Insulin Pump Recall" letter dated July, 29 20011 to all affected customers. The letter included affected product and problem. They requested anyone who received the pump to send it back to Animas and a replacement pump will be sent at no charge. For additional information contact the Customer Support Department at (877) 937-7867.

Distribution

Worldwide Distribution: Nationwide (USA) distribution including the states of CA, CO, MA, and MD; and the countries of Australia, Finland, France, Israel, Italy, Sweden and the U.K.

Quantity

380