FDA Recall Terminated

GammaMed Flexible Applicator probe; 3.2mm diameter, braced PVDF. Varian Medical Systems, Palo Alto, CA. Mfg by Varian Medical Systems, Haan GmbH. Flexible Applicator Robes are designed for brachytherapy intracavitary treatments in areas such as vagina, vagina stump or rectum.

Recall: Z-3013-2011 · Initiated July 13, 2011

Recall

Recall Number
Z-3013-2011
Event Number
59388
Firm
Varian Medical Systems, Inc. Oncology Systems
FEI Number
2916710
Product Code
JAQ
Status
Terminated
Root Cause
Process control
Initiated
July 13, 2011
Posted
August 12, 2011
Terminated
June 20, 2012
Address
911 Hansen Way, Palo Alto, CA, 94304-1028

Description

GammaMed Flexible Applicator probe; 3.2mm diameter, braced PVDF. Varian Medical Systems, Palo Alto, CA. Mfg by Varian Medical Systems, Haan GmbH. Flexible Applicator Robes are designed for brachytherapy intracavitary treatments in areas such as vagina, vagina stump or rectum.

Reason

Certain GammaMed Flexible applicator probes may not have been heat annealed during manufacturing to prevent shrinkage after autoclave.

Action

Varian Medical Systems, Inc. sent out Customer Technical bulletins in August 2008 and Urgent Medical Device Correction/Urgent Field Safety Notices on July 13, 2011 to all affected customers. The letter included description of device and problem. The current notice supplements and modifies previous instructions and ask that they stop using any flexible applicator probes in inventory. They are to be returned to Varian Brachy Therapy for replacement. Letter go on to list possible scenarios if the length deviation is not detected and advised all personnel working in the radiotherapy department should be notified of the recall. For additional information contact (800) 360-7909.

Distribution

Worldwide Distribution: Nationwide (USA) and including the countries of France, Germany, India, Trinidad and Tobago, United Kingdom, Canada, Nepal, Ukraine, Czech Republic, Bangladesh, Thailand, Turkey, Brazil, Estonia, Switzerland, Russia, Greece, Philippines, Spain, Belgium, Kuwait, Romania, Morocco, Mexico, Columbia, Austria, Poland, Kazahstan, Vietnam, Cyprus, New Zealand, Dominican Republic, Norway, Uzebestan, Hungary, Lebanon, South Africa, South Korea, Slovenia, Venezuela, Serbia, Tunisia, Pakistan, Macedonia, Ireland, Azerbajian, Ukraine, Morroco, Italy, Singapore, Nambia, Turkey, Sudan, and Taiwan.

Quantity

1343