Eclipse Treatment Planning System; Varian Medical Systems, Palo Alto, CA Product Usage: Eclipse Treatment Planning System is used to plan radiotherapy treatments for patients with malignant or benign diseases. It is used to plan external beam irradiation with photon, electron and proton beams, as well as for internal irradiation (Brachytherapy) treatments.

Recall: Z-3002-2011 · Initiated June 30, 2011

Recall

Recall Number
Z-3002-2011
Event Number
59378
Firm
Varian Medical Systems, Inc. Oncology Systems
FEI Number
2916710
Product Code
MUJ
Status
Terminated
Root Cause
Software design
Initiated
June 30, 2011
Posted
August 11, 2011
Terminated
May 9, 2014
Address
911 Hansen Way, Palo Alto, CA, 94304-1028

Description

Eclipse Treatment Planning System; Varian Medical Systems, Palo Alto, CA Product Usage: Eclipse Treatment Planning System is used to plan radiotherapy treatments for patients with malignant or benign diseases. It is used to plan external beam irradiation with photon, electron and proton beams, as well as for internal irradiation (Brachytherapy) treatments.

Reason

In affected Eclipse versions, the treatment field energy may change from the current selection to a previous selection without use intervention and without invalidating the MU or calculated dose distribution.

Action

Varian sent an "Urgent Medical Device Correction/Urgent Field Safety Notice" letter dated June 30, 2011 to all affected consignees. The letter identifies the affected product, problem, and actions to be taken. The letter states that Varian will contact all affected customers to apply a database configuration script to the database and is preparing a correction for Eclipse Version 10 to address the issue. A Varian representative will contact each site to schedule correction when it is available. The letter also provides RECOMMENDED USER ACTIONS to follow. The Letters go on to reiterate caution and warnings from Eclipse instructions for use. All appropriate personnel should be notified of the correction notice. For future reference, this document will be posted to the Varian customer support website: http://www.MyVarian.com. For further clarification, contact your local Varian Customer Support District or Regional Manager.

Distribution

Worldwide Distribution.

Quantity

2436