ATLANTIS(R) Anterior Cervical Plate System Screw Caddy; Self Drilling Fixed Screw Caddy (976-516). Medtronic Sofamor Danek USA, Inc. This instrument is a precision device which incorporates a measuring function.
Recall
- Recall Number
- Z-2985-2011
- Event Number
- 59221
- Firm
- Medtronic Sofamor Danek USA Inc
- FEI Number
- 1000477302
- Product Code
- XLH
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- July 6, 2011
- Posted
- August 10, 2011
- Terminated
- April 26, 2013
- Address
- 1800 Pyramid Place, Memphis, TN, 38132
Description
ATLANTIS(R) Anterior Cervical Plate System Screw Caddy; Self Drilling Fixed Screw Caddy (976-516). Medtronic Sofamor Danek USA, Inc. This instrument is a precision device which incorporates a measuring function.
Lines on the screw gauge may not placed on the caddy properly which could result in improper measurement of screws.
Medtronic Sofamor Danek USA, Inc. sent "Voluntary Recall - Medtronic" letters on July, 2011 to all affected customers. The letter included; description of product and problem, and asked customers to stop using the product. Customers are asked to contact their Sales Representatives for replacement. For question on this recall call Medtronics Global Quality Department at (800) 876-3133.
Worldwide Distribution: Nationwide distribution including the states of MN, OK, IL, AK, CA, PA, UT, LA, SC, NY, MI, GA, WI, KY, AL, MO, AZ, TN, TX, MA, NE, OH, FL, CO, MT, CT, VA, NC, NH, WA, WV, WY, IN, IA, KS, MT, ME, DE, DC, NJ, MD, MS, NV, HI, ID, ND, and AR; and the country of Uruguay.
191 units