FDA Recall Open, Classified

Pneupac paraPAC plus 300 ventilator kit, REF P300NXX

Recall: Z-2973-2024 · Initiated May 31, 2024

Recall

Recall Number
Z-2973-2024
Event Number
95025
Firm
Smiths Medical ASD Inc.
FEI Number
3012307300
Product Code
BTL
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
May 31, 2024
Posted
September 11, 2024
Address
6000 Nathan Ln N, Minneapolis, MN, 55442-1690

Description

Pneupac paraPAC plus 300 ventilator kit, REF P300NXX

Reason

There is a potential for inadvertent tidal volume knob movement from the original setting when set at high (1000-1500 mL) and low (70-150 mL) settings.

Action

Smiths Medical issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 05/31/2024 by letter delivered via traceable means. The notice explained the issue, potential risk, and requested the following: "Customer Required Actions: 1. Identify all affected paraPac plus serial numbers in your possession. 2. Perform an inspection to determine if your devices are affected per the instructions below: a. Rotate the tidal volume knob to the lowest position of the control and evaluate whether or not the knob stays in position as set or if it moves to a higher position. b. Repeat this step, rotate tidal volume knob to the highest position of the control and evaluate whether the knob stays in position as set or if it moves to a lower position. c. If the knob stays in the position as set when tested, then your product is not affected, and you can continue use of the device as normal. Complete the attached Response Form confirming that you have no affected product. d. If the knob moves or changes position from the set position, then your product is affected, and it will need to be removed from use and repaired. Do not attempt to use or repair the affected product. Report the event to Global Complaint Management at [email protected]. Complete the attached Response Form identifying the number of affected products in your possession. 3. Share this recall notification with all potential users of the devices to ensure they are aware of this recall and proposed mitigations. 4. Return the attached Customer Response Form as noted above to [email protected] within ten days of receipt to acknowledge your understanding of this notification. 5. DISTRIBUTORS: If you have distributed potentially affected products to your customers, please immediately forward this notice to them. Request that they complete the response form and return it to [email protected]." Questions: https://icumed.custhelp.com/app/market- action; Technical

Distribution

Worldwide

Quantity

2198 units