FDA Recall Terminated

Terumo Advanced Perfusion System 1 Ultrasonic Air Sensor 1/4x1/16 UAS BLK, Catalog Number 5785 - Product Usage: intended for use with Terumo and Sarns brand air detection and heart-lung systems to detect gross air bubbles in the line during extracorporeal procedures.

Recall: Z-2960-2020 · Initiated August 25, 2020

Recall

Recall Number
Z-2960-2020
Event Number
86338
Firm
Terumo Cardiovascular Systems Corporation
FEI Number
1828100
Product Code
DTQ
Status
Terminated
Root Cause
Employee error
Initiated
August 25, 2020
Terminated
January 29, 2021
Address
6200 Jackson Rd, Ann Arbor, MI, 48103-9586

Description

Terumo Advanced Perfusion System 1 Ultrasonic Air Sensor 1/4x1/16 UAS BLK, Catalog Number 5785 - Product Usage: intended for use with Terumo and Sarns brand air detection and heart-lung systems to detect gross air bubbles in the line during extracorporeal procedures.

Reason

Internal testing at the supplier, revealed that a crimped wire was placed into an air bubble detector (ABD) sensor connector backwards, preventing the crimped wire from locking into the mating wire connector. This intermittent connection may result in false air bubble alarms.

Action

Urgent Medical Device Removal notification letters dated 8/25/20 were sent to customers. Terumo CVS is removing the affected ABD sensors from the field. A Terumo Associate will contact all affected customers to coordinate the shipment of a replacement device and subsequent return of the affected unit. The replacement ABD sensor supplied by Terumo CVS will be part of a kit containing a cable and a bracket which are not affected by this removal. The cable and the bracket do not require replacement. CUSTOMER INSTRUCTIONS 1. Review this Medical Device Correction and assure that all users have received notice of this issue. 2. Confirm receipt of this communication by completing and returning the attached Customer Response Form as indicated on the form. 3. Review Product Return instructions when provided by separate email communication. Tracking information on replacement device will be provided as well as a Return Goods Authorization (RGA) number to facilitate the return of the affected device., 4. Once the replacement shipment has been received, remove the affected ABD sensor and replace it with the new ABD sensor provided by Terumo CVS. 5. Return affected ABD sensor using the packing container and shipping materials received with the replacement ABD device. Please mark the RGA number on the outside of the box. QUESTIONS? We encourage you to contact Terumo CVS with any questions or concerns: Terumo CVS Customer Service: 1.800.521.2818 Monday Friday, 8 a.m. 6 p.m. ET Recall Fax: 1.734.741.6149 Review this Medical Device Correction and assure that all users have received notice of this issue. A Terumo Associate will contact all affected customers to coordinate the shipment of a replacement ABD sensor and subsequent return of the affected unit. Terumo CVS recommends that you continue using your ABD sensor while waiting for a replacement. If you experience an ABD sensor failure, or have questions, contact Terumo CVS Customer Service: 800.521

Distribution

Worldwide distribution - US Nationwide distribution.

Quantity

3