FDA Recall Terminated

Overhead Counterpoise Systems (OCS), OCS CONF,STD CEIL MNT, STD HORZ ARM, Catalog # OCS 115A, Material Number 59880253

Recall: Z-2960-2018 · Initiated June 28, 2018

Recall

Recall Number
Z-2960-2018
Event Number
80727
Firm
Bayer Medical Care, Inc.
FEI Number
2520313
Product Code
JAK
Status
Terminated
Root Cause
Equipment maintenance
Initiated
June 28, 2018
Terminated
November 26, 2018
Address
1 Bayer Dr, Indianola, PA, 15051-9702

Description

Overhead Counterpoise Systems (OCS), OCS CONF,STD CEIL MNT, STD HORZ ARM, Catalog # OCS 115A, Material Number 59880253

Reason

The torque wrench used on the ceiling column assembly during a recent service visit was outside of the torque specification, potentially overtightening the bolts. The overtightening of bolts may compromise the structural integrity of the system which could lead to the OCS disengaging from the ceiling mount and this may cause injury to the patient or the operator.

Action

On June 28, 2018, Urgent Medical Device Field Safety Corrective Action letters were distributed to customers. The letters informed the customers of the following: A Bayer Service Representative will be scheduling a service visit to replace the ceiling column assembly of your OCS system(s) at no cost to you.

Distribution

The devices were distributed to the following US states: OH, TX, and WV.

Quantity

2