FDA Recall Terminated

GE Carestation 30 anesthesia systems

Recall: Z-2955-2018 · Initiated July 24, 2018

Recall

Recall Number
Z-2955-2018
Event Number
80693
Firm
GE Medical Systems China Co., Ltd. National Hi-Tech No.
FEI Number
3003513129
Product Code
BSZ
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
July 24, 2018
Terminated
December 16, 2020
Address
19 Changjiang Road, Dev. Zone Wuxi China

Description

GE Carestation 30 anesthesia systems

Reason

On systems equipped with the Auxiliary Common Gas Outlet (ACGO), there is a potential for the auxiliary gas switch to unexpectedly switch from the ACGO position back to the circle position. On systems equipped with the Aux O2 + Air function the switch could unexpectedly switch from the Aux O2 + Air position back to the circle position. These situations could lead to hypoxia.

Action

The firm issued a medical device correction notice on 07/24/2018. The notice explained the issue and provided safety instructions should the use observe a switch change. The units will be corrected. Questions or concerns may contact 1-800-437-1171 or the local service representative.

Distribution

Worldwide

Quantity

252 units