GE Carestation 620/650/650c, A1 and A2 (non-US distribution) anesthesia systems

Recall: Z-2954-2018 · Initiated July 24, 2018

Recall

Recall Number
Z-2954-2018
Event Number
80693
Firm
GE Medical Systems China Co., Ltd. National Hi-Tech No.
FEI Number
3003513129
Product Code
BSZ
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
July 24, 2018
Terminated
December 16, 2020
Address
19 Changjiang Road, Dev. Zone Wuxi China

Description

GE Carestation 620/650/650c, A1 and A2 (non-US distribution) anesthesia systems

Reason

On systems equipped with the Auxiliary Common Gas Outlet (ACGO), there is a potential for the auxiliary gas switch to unexpectedly switch from the ACGO position back to the circle position. On systems equipped with the Aux O2 + Air function the switch could unexpectedly switch from the Aux O2 + Air position back to the circle position. These situations could lead to hypoxia.

Action

The firm issued a medical device correction notice on 07/24/2018. The notice explained the issue and provided safety instructions should the use observe a switch change. The units will be corrected. Questions or concerns may contact 1-800-437-1171 or the local service representative.

Distribution

Worldwide

Quantity

310 units