FDA Recall Terminated

Hemochron JR. Microcoagulation Systems Prothrombin Time (PT). International Ethidine Corp (ITC)-Nexus Dx. A unitized mlcrocoagulation test intended for use in performing one-stage prothrombin time.

Recall: Z-2953-2011 · Initiated May 10, 2011

Recall

Recall Number
Z-2953-2011
Event Number
58043
Firm
International Technidyne Corp.
FEI Number
1000526865
Product Code
JPA
Status
Terminated
Root Cause
Device Design
Initiated
May 10, 2011
Posted
August 5, 2011
Terminated
November 2, 2012
Address
68 Olsen Ave, Edison, NJ, 08820-2419

Description

Hemochron JR. Microcoagulation Systems Prothrombin Time (PT). International Ethidine Corp (ITC)-Nexus Dx. A unitized mlcrocoagulation test intended for use in performing one-stage prothrombin time.

Reason

Certain lots of Hemochron Jr. PT Cuvettes (J201) demonstrate a higher bias than historically observed when compared to laboratory reference instruments.

Action

International Technidyne Corp. (ITC) sent "Urgent Medical Device Recall" letters dated May 10, May 25 and June 09, 2011 to all affected customers. The letter included a description of the problem, asked customers fill out the attached form, and return affected product for replacement. For questions, please contact ITC Technical Support at (800) 631-5945.

Distribution

Worldwide Distribution: (USA) Nationwide Distribution including the countries of; Austria, Australia, Canada, Switzerland, Germany, Denmark, Estonia, Spain, Finland, France, United Kingdom, Hong Kong, Ireland, Israel, India, Italy, Korea, Liechtenstien, Libya, Netherlands, Norway, Oman,Portugal, Qatar, Saudi Arabia, Sweden, Turkey, Taiwan, and South Africa.

Quantity

6,218 boxes in US ; 4713 boxes outside US