Hemochron JR. Microcoagulation Systems Prothrombin Time (PT). International Ethidine Corp (ITC)-Nexus Dx. A unitized mlcrocoagulation test intended for use in performing one-stage prothrombin time.
Recall
- Recall Number
- Z-2953-2011
- Event Number
- 58043
- Firm
- International Technidyne Corp.
- FEI Number
- 1000526865
- Product Code
- JPA
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- May 10, 2011
- Posted
- August 5, 2011
- Terminated
- November 2, 2012
- Address
- 68 Olsen Ave, Edison, NJ, 08820-2419
Description
Hemochron JR. Microcoagulation Systems Prothrombin Time (PT). International Ethidine Corp (ITC)-Nexus Dx. A unitized mlcrocoagulation test intended for use in performing one-stage prothrombin time.
Certain lots of Hemochron Jr. PT Cuvettes (J201) demonstrate a higher bias than historically observed when compared to laboratory reference instruments.
International Technidyne Corp. (ITC) sent "Urgent Medical Device Recall" letters dated May 10, May 25 and June 09, 2011 to all affected customers. The letter included a description of the problem, asked customers fill out the attached form, and return affected product for replacement. For questions, please contact ITC Technical Support at (800) 631-5945.
Worldwide Distribution: (USA) Nationwide Distribution including the countries of; Austria, Australia, Canada, Switzerland, Germany, Denmark, Estonia, Spain, Finland, France, United Kingdom, Hong Kong, Ireland, Israel, India, Italy, Korea, Liechtenstien, Libya, Netherlands, Norway, Oman,Portugal, Qatar, Saudi Arabia, Sweden, Turkey, Taiwan, and South Africa.
6,218 boxes in US ; 4713 boxes outside US