FDA Recall Terminated

BrightView X Gamma Camera, 3/8". Philips Medical Systems, Inc. Intended to produce images depicting the anatomical distributions of single photon emitting radioisotopes within the human body for interpretation by medical personnel.

Recall: Z-2947-2011 · Initiated June 15, 2011

Recall

Recall Number
Z-2947-2011
Event Number
59252
Firm
Philips Medical Systems
FEI Number
1525965
Product Code
KPS
Status
Terminated
Root Cause
Software design
Initiated
June 15, 2011
Posted
August 4, 2011
Terminated
February 15, 2013
Address
3860 N 1st St, San Jose, CA, 95134-1702

Description

BrightView X Gamma Camera, 3/8". Philips Medical Systems, Inc. Intended to produce images depicting the anatomical distributions of single photon emitting radioisotopes within the human body for interpretation by medical personnel.

Reason

Non-conformances with BrightView X/XCT systems running version 2.0.1 including data transfer, XCT acquisition, Pixel Mask for image reconstruction and table moving incorrectly.

Action

Philips Medical Systems, Inc. set a "Customer Information" letter dated July 8, 2011 to all affected customers. The letter informed them of an upcoming software updated (version 2.5). For questions please contact 800-722-9377 (option 5).

Distribution

Worldwide Distribution: (USA) Nationwide distribution and the countries of Greece, Australia, Belgium, China, Czech Republic, France, Germany, Indonesia, Israel, Japan, Lithuania, Malaysia, Netherlands, New Zealand, Norway, Poland, Portugal, Puerto Rico, Saudi Arabia, Slovakia, South Korea, Spain, Switzerland, Taiwan U.A.E., and the UK.

Quantity

213 units, all systems