FDA Recall Open, Classified

Trilogy Evo Universal Ventilator, Model No. DS2000X11B

Recall: Z-2943-2024 · Initiated July 16, 2024

Recall

Recall Number
Z-2943-2024
Event Number
95078
Firm
Philips Respironics, Inc.
FEI Number
2518422
Product Code
CBK
Status
Open, Classified
Root Cause
Device Design
Initiated
July 16, 2024
Posted
September 5, 2024
Address
1001 Murry Ridge Ln, Murrysville, PA, 15668-8517

Description

Trilogy Evo Universal Ventilator, Model No. DS2000X11B

Reason

Firm has released a mandatory software update Version 1.05.10.00 and a User Manual addendum for affected ventilators. These updates resolve safety issues identified in earlier recalls.

Action

On July 16, 2024, the firm began notifying customers through letters titled "URGENT Medical Device Correction" - UPDATE. Customers were informed of the release of mandatory software update version 1.05.10.00 and a User Manual addendum for the Trilogy Evo, Trilogy Evo O2, Trilogy Evo Universal, and Trilogy EV300 ventilators. This letter outlines the safety issues addressed by this update, previously tracked through recalls RES 91832, 92048, 92515, 93609, and 94263. Issues include "Accuracy of Oxygen Delivery", "Environmental Contamination of Device Sensor", "Battery Depleted or Loss of Power Alarms Triggered", "Incomplete Contraindications Statement", and "Translation errors in the Korean, Traditional Chinese, and Spanish Instructions for Use manuals". Customers were instructed to immediately update the device software following instructions provided in the letter. The firm will follow up with customers to ensure devices have been updated. DME/Homecare users can access the software via the "My Philips for Professionals" website. Hospital users can access the software via In Center. If you have any questions, call 800-345-6443.

Distribution

Worldwide distribution: US (nationwide) and OUS (foreign) countries of: AU, BD, BE, BG, BO, BR, BS, CH, CL, NL, CN, CO, CZ, DE, DK, FR, SE, GB, GF, GR, GT, HK, HR, HU, IE, IL, IN, IT, JP, KE, KW, LB, LT, MA, ME, MA, MT, MQ, MX, NO, NP, OM, PA, PE, PF, PK, PL, PT, PY, QA, RE, RO, RS, RU, SA, SE, SI, SK, SG, TH, TN, TR, TW, UA, YT, ZA and ZW.

Quantity

90,910 US; 56,273 ROW