Stryker Medical Stretcher Model 1105 Prime 5th Wheel A wheeled stretcher consists of a platform mounted on a wheeled frame that is designed to transport patients in a horizontal position. The device has siderails and supports for fluid infusion equipment.

Recall: Z-2927-2011 · Initiated July 19, 2011

Recall

Recall Number
Z-2927-2011
Event Number
59214
Firm
Stryker Medical Division of Stryker Corporation
FEI Number
1831750
Product Code
INK
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
July 19, 2011
Posted
August 3, 2011
Terminated
August 28, 2013
Address
3800 East Centre Ave., Portage, MI, 49002

Description

Stryker Medical Stretcher Model 1105 Prime 5th Wheel A wheeled stretcher consists of a platform mounted on a wheeled frame that is designed to transport patients in a horizontal position. The device has siderails and supports for fluid infusion equipment.

Reason

Stryker Medical has determined that some Prime Brake Pad Assembly components were manufactured with brake shoes that did not meet specifications. Accordingly some brake shoes used within the Prime Brake Pad Assembly may wear prematurely, over the life of the product. Reduced braking force may result in a patient fall or caregiver muscle strain, either of which may require medical intervention.

Action

Stryker Medical sent an "URGENT- MEDICAL DEVICE CORRECTION" letter dated July 18, 2011 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter instructs customers to locate the affected stretchers and to test the brakes. If the brakes are found to be inoperable, customers are instructed to remove the stretchers from service and contact the firm at 800-STRYKER, option 3. Additionally, customers are instructed not to put the stretchers back in service until they are repaired by the firm's field service representative. For units with properly functioning brakes, the letter also instructs the user to test the brakes before each use until the unit has been serviced. A post card was enclosed to confirm receipt of the letter. Questions or concerns may be addressed by calling Stryker at 269-389-6604.

Distribution

Worldwide Distribution-USA (nationwide) including the states of AL, CA, FL, GA, HI, IL, MA, MN, MO, NH, NJ, NV, NY, PA, RI, TX, UT, and VA and the country of THE NETHERLANDS.

Quantity

23